4.5.4  Setting the calibration factor (CAL) 

 

 

 
 

When the meter is normally turned on, the calibration factor set value will display for 2 seconds. 

 

 

 

 

 

 

 

 

1. Press 

 key to turn off the meter. 

2. Press and hold on 

key, then press 

 key to turn on the meter to enter the calibration 

factor setting mode, the “

CAL SEt

” marks are displayed. 

3. Press 

 or 

 key to increase or decrease the value. 

4. Press 

 key to store the new setting value and exit. 

 

4.5.5  Switching the alarm function on or off 

 

1. Press 

 

key for 3 seconds to switch the alarm function on or off. The “

ALARM

” and “

” symbols in the display indicates that the alarm function is on. 

2. 

When the alarm function is on, the display will show “

” if the measured value exceeds the 

limit. 

 

exceed limit value

on/off

ALARM

3 seconds

Summary of Contents for 593R

Page 1: ...Manual Elma 593R ELECTROSMOG METER EN EAN 5706445340330...

Page 2: ...5 4 OPERATION 6 4 1 Front panel controls description 6 4 2 LCD display description 8 4 3 Use E field sensor 9 4 4 Explanatory notes 9 4 5 Setting the meter 11 4 6 Making measurements 15 4 7 Manual da...

Page 3: ...serve the operating instructions for equipment which is used to generate conduct or consumer electromagnetic energy Be aware that secondary radiators e g reflective objects such as a metallic fence ca...

Page 4: ...magnetic flux density divided by the permeability of the medium Magnetic field strength is expressed in units of amperes per meter A m The meter uses of magnetic field strength for measurements only...

Page 5: ...the range from 10MHz to 8GHz The non directional electric field and high sensitivity also allow measurements of electric field strength in TEM cells and absorber rooms The unit of measurement and the...

Page 6: ...Operating temperature range 0 to 50 Operating humidity range 25 to 75 RH Storage temperature range 10 to 60 Storage humidity range 0 to 80 RH Dimensions Approx 67 W 60 T 247 L mm Weight including bat...

Page 7: ...us Max instantaneous Max average 5 key Press this key to change unit s selector mV m or V m A m or mA m W m2 mW m2 or W m2 W cm2 or mW cm2 Press and hold this key while turning the meter on to disable...

Page 8: ...device to the calibration factor setting mode press key to exit this mode 9 key Press this key one time to store one data set to memory Press and hold this key while turning the meter on to enter the...

Page 9: ...he memory 9 Read manual data memory mode indication 10 Manual data memory address number 1 99 The manual data memory full indication 11 ALARM Alarm function on off or alarm setting indication 12 When...

Page 10: ...ing to be moved in the 3 planes Simply point it at the target to make the measurement 4 4 Explanatory notes 4 4 1 Units of measurement The meter measures the electrical component of the field the defa...

Page 11: ...ded for this application In many cases the measurement accuracy will be sufficient even if the frequency response of sensor calibration factor is ignored CAL can be set to 1 00 in such cases E Field t...

Page 12: ...the units of measurement With the key as follows a Electric field strength V m b Computed magnetic field strength mA m c Computed power density mW m2 d Computed power density W cm2 X Y Z Z Y X 2 X Y Z...

Page 13: ...Elma 593R 12 4 5 2 Setting the result mode Instantaneous result mode is automatically set when the meter is turned on With the key as followings MAX AVG MAX Instantaneous Max Instantaneous Max average...

Page 14: ...turn on the meter to enter the alarm setting mode the mW m2 unit is flashing displayed 3 Press and keys to select the desired setting unit 4 Press key to enter the alarm value setting mode the one of...

Page 15: ...tor setting mode the CAL SEt marks are displayed 3 Press or key to increase or decrease the value 4 Press key to store the new setting value and exit 4 5 5 Switching the alarm function on or off 1 Pre...

Page 16: ...urned on the auto power off function is on 1 Press key to turn off the meter 2 Press and hold key and turn on the meter again to disable the auto power off function the symbol will disappear from the...

Page 17: ...radiation Make measurements at those points where parts of your body are nearest to the source of radiation Note Use the Max average modes only when the instantaneous measurement values are fluctuati...

Page 18: ...ss key again to exit 4 7 3 Deleting manual data memory measured values Once the memory is full you can clear the entire contents of the manual data memory 1 Press to turn off the meter 2 Press and hol...

Page 19: ...cklight auto off 5 BATTERY CHARGING 1 Battery capacity Indication Battery Capacity Indication Battery Capacity 100 capacity 40 capacity 80 capacity 20 capacity 60 capacity 5 2 When the battery capacit...

Page 20: ...v 210823 23 08 2021 14 34 31 Z Y X AVG MAX 2...