7. User’s maintenance
Warning:
Please comply with sterilization regulations and security disciplines because all
used equipments possibly contain blood and body fluid of the animal.
Warning:
Personnel without maintenance experience to this kind of equipments are
prohibited undertaking maintenance tasks.
7.1Maintenance schedule
Least maintenance
times
Maintenance
Every day
Clean external surface
21%O
2
calibration
Every two weeks
Empty evaporator, and discard remanent agent
Every month
100%O
2
calibration
During cleaning and
installation
Check whether accessories are in good condition or
not, and replace or service them when necessary
Process according to
need
Empty water store, and replace absorbent in
absorber
7.2Maintenance of anaesthetic evaporator
Anaesthetic evaporator is the most important part of the workstation, therefore notice to:
never lift the concentration-regulating knob by hand; in case of regulating the anaesthetic
concentration one must firstly gently rotate the dial gauge of knob, and mustn't further
rotate the gauge forcibly when reaching already the limit. Suggest calibrating the
evaporator annually. Oxygen should be used as the carrier when checking the evaporator.
The flowrate pass through the evaporator are 1
±
0.1L/min and 5
±
0.5L/min.