SAFETY INFORMATION
Accessory equipment connected to the analog and digital interfaces must be certificated
according to the respective IEC/EN standards (e.g. IEC/EN 60950 for data processing
equipment and IEC/EN 60601-1 for medical equipment). Furthermore all configurations shall
comply with the system standard EN 60601-1-2:2007. If you connect additional equipment to
the signal input or signal output of this device, then you are configuring a medical system, and
are therefore responsible that the system complies with the requirements of the system
standard EN 60601-1-1:2001. If in doubt, consult the technical service department or your local
representative.
For EU Countries
The following mark, the name & address of the EU Representative shows compliance of the
instrument with Directive 93/42/EEC
.
ISO 15004
This report provides information about the hazard to the examinee
’s eyed in compliance
with ISO 15004-1:2006, ISO 15004-2:2007 Ophthalmic instruments
– Fundamental
requirements and test methods Part2
– Light hazard protection.
This condition is satisfied
even when the instrument is operating at maximum light intensity and maximum aperture!
(Maximum intensity is the highest brightness the instrument is capable of delivering,
including the highest brightness achievable if overvoltage is provided) detailed radiation
information at normal usage of this instrument is like bellows.
Radiation output: below 117.1
μW/cm2 Limit by ISO15004: 100 mW/cm2
Number
Radiation output [μW/cm2]
1
107.0
2
117.1
3
115.5
4
115.7
5
103.6
6
103.7
7
108.8
8
109.0
9
105.6
10
105.8
average
109.1