Installation and First Start-Up
- 13 -
7.2
Mounting the Evaluation Box
The evaluation box has two lateral Ø 4.5 mm mounting holes for mounting the housing on a surface with two
M4 screws. The drilling distance is 106 mm.
see also section
6.3 Dimensions of Evaluation Box
7.3
Mounting the Sensor Head
To mount the sensor head, use 2 x M3 screws of adequate length.
see also section
During installation, the specified installation tolerances must also be adhered to:
7.3.1
Mounting Tolerances
Table1: Mounting tolerances
Magnetic tape MB20-50-10-1-R
Sensor distance to the tape: max. 1.5 mm
Lateral offset: ± 3 mm
Tilt and pitch angle: ± 5°
Yaw angle: ± 1.5°
Figure 6: Mounting tolerances
max. 1.5 mm
± 3 mm
± 1.5°
± 5°
Sensor distance:
Lateral offset:
Pitch angle:
Tilt angle:
Yaw angle: