PURELAB 7000 - US Operator Manual
ELGA
Page 14
PURELAB 7000 - US MANU40676 Version 02 - 02/16
3.3
Setting Consumable Replacement Reminders
Step 1 - Enter consumable replacement timer set-up
1.
ENSURE process is OFF.
2.
PRESENT Master Key and PRESS
button to
enter the Consumable Timer set-up.
Step 2 - Labpure L1 replacement date
1.
PRESS
button to reset cartridge replacement
date
OR
PRESS
✔
to accept replacement date and
proceed to vent filter replacement.
2.
PRESS
✔
to confirm that resetting is required
OR
PRESS
✗
to abort reset.
3.
PRESS
✔
.
Step 3
– Composite vent filter replacement date
1.
PRESS
button to reset filter replacement date
OR
PRESS
✔
to accept date and proceed to UV lamp
replacement.
2.
PRESS
✔
to confirm that resetting is required
OR
PRESS
✗
to abort reset.
3.
PRESS
✔
.
Step 4
– UV lamp replacement date
1.
PRESS
button to reset UV lamp replacement
date
OR
PRESS
✔
to accept replacement date and
proceed to Protek L1 / L2 replacement.
2.
PRESS
✔
confirm that resetting is required
OR
PRESS
✗
to abort reset.
3.
PRESS
✔
.
Labpure L1 screen
Composite vent filter screen
UV lamp screen