Page 12 of 20 M6847 issue 1.1
8. TECHNICAL SPECIFICATION.
8.1. Specifications.
Mains Power Supply
Voltage
MH8517
230V – AC 50/60 Hz or
MH8517x1
115V – AC 50/60 Hz as stated on the data plate
label.
Maximum Power
Consumption
240 Watt max at 230V
240 Watt max at 115V
Operational temperature
and performance.
Temperature stability
70
°
C nominal from ambient (25°C, with 2.3 litres water)
obtained within 70 minutes. Or 45°C within 40 minutes.
±0.5°C at constant ambient temperature, improved to ±0.25°C
when using the lid AZ9241.
Mains Power Lead set
(UK) 13A BS1362
3 core earthed / ground. 2 meters long
Moulded IEC plug and Lead set – supply cord H05 V V-F-
Replace only with equivalent cable.
Mains Power Lead set
(Europe)
3 core earthed / ground. 2 meters long
Moulded IEC plug and Lead set – supply cord H05 V V-F-
Replace only with equivalent cable.
Mains Power Lead set
(USA)
3 core earthed / ground. 2 meters long
Moulded IEC plug and Lead set – supply cord SJT VW 1-
Replace only with equivalent cable.
Lead set plug fuse (UK –
only)
13A BS1362
Product Fuse Rating
230V – F1.25A (IEC127) Quick Blow Fuse 20mm x 5mm
Glass.
115V – F5A (IEC127) Quick Blow Fuse 20mm x 5mm Glass.
Heating Element
Silicone Rubber Mat Heater.
Over Temperature
Protection
Over temperature protection with resettable thermal cut out.
Maximum Bath Capacity
2.3 litres (0.6 US gallon), (0.5 UK gallon).
Case Construction
Aluminium bath with PTFE coated interior. Powder coated
aluminium white exterior.
Neon indicators
Power on – Clear
Heater on – Amber
Controls: Bath Heater
Electronic temperature controller with thermocouple sensor
feedback
The Ingress protection rating for this product is classified as IPX 1.
8.2. Dimensions and Weight.
(unpacked and empty)
Weight 2.7Kg