Set 4-20mA Maximum (20mA) Range
UuuMAXmmm.mm
Where
uu = unit number
mmm.mm = level for 20mA output in inches (Default is 240.00”)
Example: For 20-foot long sensors = 240.00
Response:
UuuMAXOK
Set the Level Error setting (version 3.09 and higher)
UuuSETERRx
Where
uu = unit number
x = 0 will set the level error report to be 999.99. This is the default setting.
x = 1 will set the level error report to be 000.00.
Response: UuuSETERROK
Set the Modbus 16 bit Unsigned Integer, 32 bit or 2 x 16 bit floating point mode
UuuIFxxxx
Where
uu = unit number
xxxx = 1007 will set the 16 bit Unsigned integer mode. This is the default setting.
xxxx = 1008 will set the 32 bit floating point mode.
xxxx = 1009 will set the 2x16 bit floating point mode (
v3.14 and higher
).
Response: UuuIFOK
Force sensor to enter Boot Load mode (version 3.15 and higher, (future use)
U
UU
FB
uu = two-digit unit number
Response:
none
Enter High Level Electronic Shut Down (ESD mode) (version 3.15 and higher)
UuuESDONn
uu = unit number
n = one-digit number from 1 to 3 which represent the number of level request commands that will respond
with the maximum level without cycling power. After this number, the sensor reverts to normal operation
(Note: If power is cycled the count reverts back to programmed number of polls).
Response:
UuuESDONOK
ESD mode was successfully activated
Exit High Level Electronic Shut Down (ESD mode) (version 3.15 and higher)
UuuESDOFF
Response:
UuuESDOFFOK
ESD mode exit successfully
Note:
UuuESDON0 also turns off ESD mode