19
S
PINDLE
SELECTION
TABLES
R
6 SPINDLE SELECTION TABLES
6.1 L and TL Spindles (For use with RV1-L / RV2-L Viscometers).
Table 1: Spindle type L
a
a.
Type L spindles are the standard spindles supplied with RV1-L/RV2-L Viscometers.
Spindle
speed (rpm)
Maximum
b
measurable viscosity (mPa•s)
b.
The minimum recommended reading is 15% of the full scale value.
L1
c
c.
The L1 spindle is used for low viscosity measurement. The spindle guard must be used with this spindle in
order to obtain correct viscosity readings. For higher accuracy when measuring low viscosity samples, use
the Adaptor for Low Viscosity Materials and the special cylindrical spindle (LCP); See “Spare parts and
accessories” on page 28.
L2
L3
L4
0.3
20 000
100 000
400 000 2 000 000
0.5
12 000
60 000
240 000 1 200 000
0.6
10 000
50 000
200 000 1 000 000
1
6000
30 000
120 000
600 000
1.5
4000
20 000
80 000
400 000
2
3000
15 000
60 000
300 000
2.5
2400
12 000
48 000
240 000
3
2000
10 000
40 000
200 000
4
1500
7500
30 000
150 000
5
1200
6000
24 000
120 000
6
1000
5000
20 000
100 000
10
600
3000
12 000
60 000
12
500
2500
10 000
50 000
20
300
1500
6000
30 000
30
200
1000
4000
20 000
50
120
600
2400
12 000
60
100
500
2000
10 000
100
60
300
1200
6000
200
30
150
600
3000
Resolution
1 mPa•s
1 mPa•s
10 mPa•s 10 mPa•s
2300_TMA_0377_00_01.book Page 19 Thursday, March 29, 2007 1:32 PM