![ELCART 09/08500-00 Operator'S Instruction Manual Download Page 6](http://html1.mh-extra.com/html/elcart/09-08500-00/09-08500-00_operators-instruction-manual_2377019006.webp)
ELCART
ELCART DISTRIBUTION SPA via Michelangelo Buonarroti, 46 - 20093 Cologno Monzese (Milano) ITALY
Tel. +39 02.25117310 Fax +39 02.25117610 sito internet: www.elcart.com e-mail: [email protected]
Manuale di istruzioni/Scheda tecnica
La divulgazione dei dati contenuti in questa scheda è da ritenersi un servizio puramente informativo e non costituisce alcun vincolo da parte della Elcart in merito a prestazioni ed utilizzo del prodotto.
The divulgation of data contained on this technical sheet are exclusively for informational reasons and establish no link on behalf of Elcart regard to thr performances and the usa of the product.
La divulgacion de los datos contenidos en esta ficha son un servicio unicamente informativo y no constituyen ningun vinculo de parte de Elcart respecto a las prestaciones y uso del producto.
ART. 09/08500-00
PAGINA 6 DI 6
4.2 AC CURRENT
Frequency Range:
50Hz to 60Hz
Response:
Average, Calibrated in rms of sine wave
Overload protection:
1200A within 60 seconds
Jaw Opening:
Ø50mm
4.3 INSULATION TEST
(With option 500V insulation tester unit)
4.4 AC VOLTAGE
Input Impedance:
≥9MΩ on all ranges
Overload protection: 1000V DC or 750V AC on all ranges
Frequency Range:
(50Hz to 400Hz) / ≤600V
(50Hz to 200Hz) / 750V
Response:
Average, calibrated in rms of sine wave
4.5 DC VOLTAGE
Input Impedance:
≥9MΩ
Overload protection:
250Vrms AC for 200mV range
1000V DC or 750V AC for other range
4.6 RESISTANCE
Overload protection: 250VDC or 250Vrms AC on all ranges
Open circuit voltage:
700mV
4.7 FREQUENCY
5. ACCESSORIES COME WITH THE CLAMP METER
Test Leads Model: T3000
Battery 9V 6F22 or equivalent
Operation Manual
6. BATTERY REPLACEMENT
If the sign “BAT” appears on the LCD display, it indicates
that battery should be replaced. Remove the battery cover
of case. Replace the exhausted battery with a new one.
WARNING
Before attempting to open the battery cover, be sure that
test leads have been disconnected from measurement
circuits to avoid electric shock hazard.
WARNING
Using this appliance in an environment with a strong radiated
radio-frequency electromagnetic field (approximately 3V/m)
may influence its measuring accuracy.
Range
Resolution
Accuracy
200A
0.1A
±2.5% of rdg ±5 digits
1000A
1A
±3.0% of rdg ±10 digits
Range Resolution
Accuracy
Note
20MΩ
10kΩ
±2% of rdg ±2 digits
2000MΩ
1MΩ
±4% of rdg ±2 digits ≤500MΩ
±5% of rdg ±2 digits ≥500MΩ
Range
Resolution
Accuracy
200V
0.1V
±1.0% of rdg ±5 digits
750V
1V
±1.2% of rdg ±5 digits
Range
Resolution
Accuracy
2kHz
1Hz
±2.0% of rdg ±5 digits
Range
Resolution
Accuracy
2V
1mV
±0.5% of rdg ±3 digits
20V
10mV
200V
0.1V
1000V
1V
±0.8% of rdg ±3 digits
Range
Resolution
Accuracy
200Ω
0.1Ω
±1.0% of rdg ±5 digits
2kΩ
1Ω
±1.0% of rdg ±8 digits
20kΩ
10Ω
200kΩ
100Ω
2MΩ
1kΩ
Made in China
User information ex art. 26 D. 49/2014
The symbol labelled on the appliance (Annex IX D. 49/2014) indicates that
the rubbish is subject to “separate collection” and it has been placed on the
Italian market after the December 31, 2010.
The user must therefore assign or (have collected) the rubbish to a
treatment facility according to indications by the local administration, or
hand it over to the reseller in exchange for an equivalent new product. The
separate collection of the rubbish and the subsequent treatment, recycling
and disposal operations encourage the production of appliances made
with recycled materials and reduce negative effects on health and the
environment caused by improper treatment of rubbish.
In the case of very small WEEE (no external dimension more than 25 cm),
the user is eligible to get free of charge assignation to retail shops with sales
areas relating to EEE of at least 400 m².
IMPORTED AND DISTRIBUTED BY:
ELCART DISTRIBUTION SPA
Via Michelangelo Buonarroti, 46
20093 COLOGNO MONZESE (MI)
ITALY
www.elcart.com - [email protected]