Calibration

This circuit requires balance adjustments for each mode (2Q/4Q) and an overall gain adjustment.

Adjust the X balance

1. Set [MODE] to [2Q].

2. Set the [BIAS] knob fully counter-clockwise.

3. Turn the [X LEVEL] knob fully counter-clockwise.

4. Turn the [Y LEVEL] knob fully clockwise.

5. Set [Y IN] to ’AC’

6. Connect a 200Hz 10VAC triangle wave to [Y IN].

7. Adjust P202 for minimum output.

Determine the bias threshold

1. Set [MODE] to [2Q].

2. Turn the [Y LEVEL] knob fully counter-clockwise.

3. Turn the [X LEVEL] knob fully clockwise.

4. Connect a 200Hz 10VAC triangle wave to [X IN].

5. Adjust the [BIAS] knob to just below the point where the VCA starts to turn on. This is called

the "threshold point".

Adjust the Y balance

1. Leave the [BIAS] control at the "threshold point".

2. Set [MODE] to [4Q].

3. Turn the [Y LEVEL] knob fully counter-clockwise.

4. Turn the [X LEVEL] knob fully clockwise.

5. Set [X IN] to ‘AC’

6. Connect a 200Hz 10VAC triangle wave to [X IN].

7. Adjust P203 for minimum output.

Adjust the gain

1. Leave the [BIAS] control at the "threshold point".

2. Set [MODE] to [2Q].

3. Set [X IN] to ‘AC’

4. Set [Y IN] to ‘DC’

5. Turn the [X LEVEL] knob fully clockwise.

6. Turn the [Y LEVEL] knob full clockwise.

7. Connect a 200Hz 10VAC triangle wave to [X IN]

8. Connect 5VDC to [Y IN]

9. Adjust P204 for an output of 10VAC.