Tare 

 

Pressing this key will set the weight/count to zero. This is useful if you have to  take the weight with the container. 
Put the empty container and press 

Tare

, the scale will zero itself now you can put the items to count/weigh and 

the scale will only display the weight/count of the items excluding the container. 

“T”

 in a circle is displayed on the 

lower left corner of the LCD when scale is in tare mode. Pressing the Tare key again will clear the tare effect. 
 

Hold Display 

 

Press the hold key when 

there is nothing

 on the platform . Now put the object on the platform. When the weight of 

the object is stable, scale will display the weight and hold the display even if you remove the weight. 

 

To un-hold 

and go to normal mode simply press the

 Hold 

key. When the scale is in Hold mode the weight display will blink. 

The hold option is good objects of bigger area than the scale’s platform area, in which case the user can’t see 
the LCD when the object is placed on the platform. 
 

CE 

 
In the user input mode, this key is used to clear the entry. If not in the user input mode, pressing this key will only 
display the weight. 

 

CONNECTING EXCEED W SCALE TO COMPUTER 

 

1.  Connect  your Exceed W scale to your computer’s available serial port using its RS232 cable (this 

interface cable is optional and is sold separately from the scale). Refer to latter part of this manual on 
how to purchase this interface cable. 

2.  You may download the sample PC program that displays the weight of items placed on Exceed W Scale. 

You may download this at: 

http://www.postagescale.com/MSI/ExceedW_PC_Program_XMRev1.zip

 

3.  Run the program and choose the correct computer serial port where your Exceed W scale is connected. 
4.  The weight displayed on your scale’s LCD is also displayed in this program. 

 

 

USER CALIBRATION OF THE EXCEED W SCALE 

 

In case the scale loses its accuracy, then  user calibration of the scale may be needed.  The load cell of this 
precision scale is normally protected against overloaded weight applied slowly, but overload caused by dropping 
the scale or d ropping something heavy on the scale may require re-calibration. 

For user calibration you may need a calibration weight. But this cal weight should be greater than ¼ of the scale 
maximum capacity. For example, in this scale of 500g,  you can take any weight greater than 125g to 500g. Let’s 
say you have 200g weight. 

Press Auto-CAL during the start-up of the scale. A cursor will blink.  Input 20000 from the keyboard. Display will 
change to CAL 0 in the first line and 0.00gm in the second line. Wait until the display change to CAL 1. Put 200g 
on the platform and wait until the scale enters the normal weighing mode. 

 

OPERATING TEMPERATURE 

 
The scale is designed to operate in a normal room temperature. If subjected  to cold or warm temperature, it 
should be allowed to reach the room temperature before operating. 

 
 

ERROR DISPLAY 

        

- - - - - - 

will display if the scale’s maximum capacity is exceeded. DO NOT OVERLOAD THE SCALE,  as this may 

 damage the sensor permanently.  

 

Summary of Contents for Exceed W 500

Page 1: ... and read from the memory of the scale Computer Interface It can display weight on the computer using the scale s RS232 cable optional and sold separately Online Software Update Software updates can be done via Internet using Exceed W PC Program Specifications Capacity 500g Graduation Resolution 0 1g step Min startup weight detection 0 3g Min weight detection normal 0 1g Accuracy 1 of reading 1 di...

Page 2: ...r than those in the scale keys then this option of counting can be used For example if you want to sample with 13 items press COUNT key Then press SET key A cursor will blink on the LCD Input 13 from the keyboard and press SET again Now put the 13 items on the platform and press Auto CAL Now you can start counting Next time if want to count the similar items you can just press COUNT key or 0 key a...

Page 3: ...e cable 2 You may download the sample PC program that displays the weight of items placed on Exceed W Scale You may download this at http www postagescale com MSI ExceedW_PC_Program_XMRev1 zip 3 Run the program and choose the correct computer serial port where your Exceed W scale is connected 4 The weight displayed on your scale s LCD is also displayed in this program USER CALIBRATION OF THE EXCEE...

Page 4: ...our computer This cable is optional for Exceed W scales and is sold SEPARATELY from the scale Please follow these instructions when updating your Exceed W scale software 1 Register your Exceed W scale You may register your scale at this link Please provide correct scale information during registration as this will become basis for scale software update Providing exact scale information will ensure...

Page 5: ...sers who use Exceed W scale to weigh very light items EXCEED W INTERFACE CABLE Exceed W RS232 interface cable is an optional accessory of the Exceed W scale This can be used to connect your scale to your computer and run application programs for Exceed W scale or perform scale software updates for Exceed W This cable is sold separately You may order directly from us or from our website ...