94
11. TEST PARAMETER
Fig. 11.2a: HV- burn test parameters 1
Full-electronic G1-1 HV DC-Tester
11.2 Parameters insulation resistance test ISO
On this menu the settings for the
insulation resistance test are made.
.
ISO tests are made using DC, with the test object isolated from the mains.
The insulation resistance between live parts and the housing is measured
and evaluated. The parameters are largely identical to the parameters for
the high voltage test (chapter 11.1).
Depending on which device variant you have, some of the parameters
given in the following may not be available or it may not be possible to
edit some parameters. The values entered are changed in the parameter
set currently open. The parameters in the following relate to the standard
device with 3000 V/6000 V/ 2 or 5 mA.
In case of devices with options fitted, the parameters may vary.
Test mode "HV / ISO / Time"
HV
- HV-test with DC voltage 6000 V /max. 5 mA (dependent on the
device type).
ISO
(option) - Insulation-test with DC voltage 6000 V max. 5000 MOhm
(dependent on the device type).
Time
- Waiting time step, eg. time to discharging the test object past
a DC-test.
Test voltage "0.05 ... 6.0 kV DC" -
set the test voltage for the HV-test .
Run up (on/off)
- activates the ramp-up function of the test voltage.
off-
no ramp-up function, the full test voltage is applied.
on
- ramp-up function active. The test voltage is increased linearly
from the start voltage up to the test voltage.
Ramp up voltage (0.01 .. 1.00 kV/s)
“ - d
efines the amount by
which the test voltage is increased linearly in a time unit. The
start voltage is used as the start value, the end value is the test
voltage set. After the test voltage is reached, the test time set is
started.
test mode ....
test voltage (kV)
Run-up (kV/s)
Run back (kV/s)
off
0.25
0.50
ISO
IS_500V
Fig. 11.2a: ISO test parameter 1
Summary of Contents for G1-1
Page 38: ...38 4 OPERATING MODES 4 2 2 Normal mode input step Full electronic G1 1 HV DC Tester...
Page 39: ...39 4 OPERATING MODES 4 2 3 Normal mode output step Full electronic G1 1 HV DC Tester...
Page 40: ...40 4 OPERATING MODES 4 2 4 Normal mode time step Full electronic G1 1 HV DC Tester...
Page 133: ...Expires 04 19 2016...