78
10. MENU SYSTEM
10.5.3
Fault handling Auto Test
If the device detects faults or at manual interruption (red reset S2/H2),
the test is interrupted immediately and a corresponding error message
displayed in plain text.
yes
- the faulty test step can be
repeated.
n o
- i t i s p o s s i b l e n o w :
- to execute the next test step in the
plan or
- to interrupt the test run.
yes
- executes the next test step
in the plan.
no
- interrupts the test run. The
result of the test is "fault" now
(see Fig. 10.30).
Fig. 10.29: Automatic test- test screen - result ACCEPT
After the end of the test plan the device indicates the result of the test.
do you want to continue with the
next test step oft the test plan?
alternatively the whole test plan
will be canceled.
Information
no
yes
there was an error during the
test!
shall the actually test step been
repeated?
Information
no
yes
Full-electronic G1-1 HV DC-Tester
0.00 [kV]
00.21 [mA]
006.0 [s]
2ISO_2HV
Status: test finish
No fault - Accept
: HV (DC)
: 2.00 kV
: 0.10 kV/s
: off
: off
: 0.70 kV
: 6.0 s
: off
: 0.50 mA
: 0.00 mA
mode
U
Umin
t
delay
Imax
Imin
4/4
Summary of Contents for G1-1
Page 38: ...38 4 OPERATING MODES 4 2 2 Normal mode input step Full electronic G1 1 HV DC Tester...
Page 39: ...39 4 OPERATING MODES 4 2 3 Normal mode output step Full electronic G1 1 HV DC Tester...
Page 40: ...40 4 OPERATING MODES 4 2 4 Normal mode time step Full electronic G1 1 HV DC Tester...