EL-CELL ECC-OPTO User Manual Download Page 7

 

Page 7 of 11

 

ECC_OPTO_manual.docx / 29.05.2013 

 

 

 

Assembly: 
 

i) 

By means of the optionally available 

ECC-REFLoad

press a small piece of 

lithium metal into port C so that the Li metal just appears at the upper narrow 
end of the feed-through hole. Attach the spring-loaded reference pin to the 
port. Skip this initial step if working without reference electrode. 

ii) 

Push the counter electrode piston into the cell body to the uppermost position. 

iii) 

Pull the piston slightly back, and insert the Li counter electrode (max. 10 mm 
in diameter, max. 1 mm thick) 

iv) 

Put the provided glass fiber separator on top.  

v) 

Place the WE material onto the center of the separator. The WE material can 
either be a self-supported electrode film, a single grain/ crystal, or a small 
amount of powder. 

vi) 

Cover the WE material with one of the provided (holed) current collectors. Use 
Al foil for excursions to potentials >1.5 V vs. Li/Li

+

, Cu foil for excursions to 

potentials between 0 and 3 V vs. Li/Li

+

vii) 

Insert the upper O-ring seal into the cell body, and place the optical window on 
top. Make sure that the window is in the center position. 

viii) 

Attach the cell lid and fasten the three hex socket screws. 

ix) 

Attach the WE contact pin to port A. 

x) 

Attach the touch screw from below and raise it to the uppermost position; then 
turn the touch screw back (counter-clockwise) by about 30°. 

xi) 

Fill the cell with electrolyte according to the following procedure. 

 

1.  Connect the provided transfer (fill) line to port B. 

2.  Charge a 5 mL syringe with ca. 1 mL of electrolyte. 

NOTE

: We recommend 

one-time use PP plastic syringes with low friction polysiloxane pistons. 

3.  Connect the syringe to the Luer adapter of the transfer line. 

4.  Pull the syringe piston back to evacuate the cell. Hold the piston a few 

seconds in the strained position. 

5.  Release the piston so that the electrolyte is being drawn deliberately into 

the cell. 

NOTE

: Never pressurize the cell during the filling procedure. 

6.  Repeat the two previous steps to complete filling. 

7.  Remove the transfer line and close port B with the provided plug.  

Optionally port B may be plugged with the second WE contact pin 
assembly. 

xii) 

Insert the cell into the stand. 

xiii) 

Connect the cell to your potentiostat/ galvanostat or battery tester via the 
2 mm banana sockets. 

 

Summary of Contents for ECC-OPTO

Page 1: ...2011 2013 EL Cell GmbH Electrochemical Test Cell ECC OPTO User Manual Release 2 0 2013 05 29...

Page 2: ...in this manual to improve reliability function or design EL Cell GmbH does not assume any liability arising from the use or application of this product Contents 1 Product Description 3 2 Features 4 3...

Page 3: ...Thus the optical instrument looks from the top through a glass window onto the backside of the WE material Typical instrumentations include optical microscopy and confocal Raman spectroscopy in the r...

Page 4: ...rains powder samples and bound electrodes self supporting or with an expanded metal holed metal foil as the current collector Maximum electrode diameter is 10 mm Easy and clean electrolyte filling via...

Page 5: ...Unpacking Check the contents of the packages against the list given below to verify that you have received all of the components Contact the factory if anything is missing or damaged NOTE Damaged ship...

Page 6: ...he cell is hermetically sealed so that it may be operated in ambient atmosphere The test cell can be used in either 2 electrode or 3 electrode reference configuration In the following the cell assembl...

Page 7: ...tical window on top Make sure that the window is in the center position viii Attach the cell lid and fasten the three hex socket screws ix Attach the WE contact pin to port A x Attach the touch screw...

Page 8: ...micals used have to be disposed properly All wetted parts are to be cleaned with water and or other appropriate solvents Ultrasonic cleaning is recommended All parts are to be dried immediately after...

Page 9: ...e 9 of 11 ECC_OPTO_manual docx 29 05 2013 7 Accessories and Spare Parts Options o ECC Opto XRD kit Beryllium window and modified lid part ECC1 00 0156 B o Sapphire window part tbd ECC Opto ECC1 00 015...

Page 10: ...repair 1 Send an e mail to info el cell com to obtain a return authorization number and a decontamination report form 2 Sign the decontamination report asserting that the instrument has been decontam...

Page 11: ...modification or the attachment of improper devices to the goods This Limited Warranty does not cover expendable items This warranty is void when repairs are performed by a non authorized person or se...