Page 9 of 17
User Manual ECC-Opto-SBS – Release 1.23
2.
The two bigger adjustment screws serve to adjust the distance and
parallelism between the two electrodes aka the width of the electrolyte
(separator) gap.
Inside the glove box, place the inner cell in front of you with the two adjustment
screws pointing to the right.
Adjust the two opposite thrust screws so as to move the rams into the lowermost
position.
Move the right ram and ram holder into the outmost right position by alternately
turning the two adjustment screws on the right. Apply only gentle torque; switch
between the two screws when torque builds up.
Place a rectangular piece of the glass fiber separator into the gap between the
squared rams. The upper edge of the separator must have the same height like
the ram holder, because the separator has to touch the glass window.