11
E.
Press the ENT button. The LCD will change to:
ENT
ENTER INTERFACE:
RBCs/PLASMA
RUN
F.
Move the slider to the RBCs/PLASMA interface. Look through the
transparent slider and position the vertical black line on the middle of
the diagonal interface as shown in the following diagram.
In some instances, a line of red blood cells extending from the
RBCs/PLASMA interface through the PLASMA/AIR interface can be
observed. These fine lines of red blood cells are residuals from the
migration and they have not been found to affect the results.
G.
Press the ENT button. The LCD will display:
ENT
ENTER INTERFACE:
PLASMA/AIR
RUN
H.
Move the slider to the PLASMA/AIR interface. Place the vertical black
line of the slider over the interface at the end of the plasma curve as
shown in the following diagram.