8. Measuring
1.
Insert the filled cuvette in the cuvette holder.
2.
Press down gently until you feel a “click” and hold in position until the result
appears on the screen. Pull the cuvette out of the DiaSpect Tm quickly.
3.
Dispose of the used cuvette in a container for potentially infectious waste.
Record the test result as soon as the checkmark is shown.
4.
The result will remain on the display until replaced by the next measurement.
To erase the latest result, press down on the empty cuvette holder.
Use only completely filled cuvettes for measuring. A filled cuvette should be
analyzed within 1 minute after filling. A filled cuvette should be kept in a horizontal
position until measurement.
If the DiaSpect Tm analyzer has been out of use for a couple of hours, an error
code may appear after the first measurement. Remove the filled cuvette, make
a “blank” measurement by pressing down the empty cuvette holder and then
reinsert the filled cuvette for measurement.