26
Chapter6Troubleshooting
Problem
Possible cause and remedy
6. The power turns off during operation and
no image is displayed.
• Whentheinternaltemperatureisextremelyhigh,the
monitorandACadapterturnoffautomatically.Ifthepower
switchisturnedonagainwhilethetemperatureremains
high,thepowerturnsoffagaininoneminute.
• Changetheinstallationlocationorenvironment,andthen
turnitonagainafewminuteslater.
-Theventilationslotmustnotbeclosed.
-Thereisnohightemperatureobjectaround.
• Ifthepowerisnotturnedonevenaftertheinstallation
locationorenvironmenthaschanged,contactyourlocal
EIZOrepresentative.
Attention
• Do not block the ventilation slots for the monitor or AC adapter,
and do not use the monitor or AC adapter in a poorly-ventilated
area.
7. The operation sound of the fan for the AC
adapter is noticeable.
• TheACadapterisequippedwithafantoavoidriseinthe
internaltemperature.Theoperationsoundmaybeheard
dependingontheinstallationlocation.However,itisnota
problem.
• ChangetheinstallationlocationoftheACadapterorthe
orientationofthefanventilationslot.
8. The Hybrid Gamma function cannot be
used.
• Checkwhethertheinputresolutionsoftheleftandright
screensarethesame.
• Thisfunctioncanbeusedonlywhentheinputresolutions
oftheleftandrightscreensarethesameorthesignals
areinputtoeitherscreenonly.
9. SelfCalibration/Grayscale Check failure
• Refertotheerrorcodetableinthe“InstructionsforUse”.
• Ifanerrorcodethatdoesnotappearintheerrorcode
tableisdisplayed,contactyourlocalEIZOrepresentative.
Attention
• Do not touch the Integrated Front Sensor.
10. Time is not displayed correctly in the
history of SelfCalibration/Grayscale
Check.
• ExecuteMonitorDetectionusingRadiCS/RadiCSLE.For
details,refertoRadiCS/RadiCSLEUser’sManual.