Alarms and Troubleshooting
244
The following sections describe the different types of alarms and messages that can be
generated by the pump, and explain how to troubleshoot common programming
issues:
.................................................................................................
..................................................................................................
Alarms Overview
The Sapphire pump generates four different types of alarms. The alarm types are
categorized according to their effect on the infusion. In all alarm types, instructions
about how to proceed (and, if relevant, to solve the problem) are displayed on the
touch screen.
When Level 1, 2, and 3 alarms occur, the red Alarm LED stays On continuously, and an
auditory alarm sounds continuously. These alarms require immediate attention.
Chapter 10: Alarms and Troubleshooting
Alarm Type
Effect on Infusion
Level 1
Pump shuts down.
Level 2
Infusion stops and cannot be reactivated.
Level 3
Infusion stops, but may be reactivated.
Message
Infusion is not interrupted.
Summary of Contents for Q Core Sapphire
Page 1: ......
Page 68: ...Sapphire Infusion Pump User Manual 67 This page is left intentionally blank ...
Page 82: ...Sapphire Infusion Pump User Manual 81 This page is left intentionally blank ...
Page 188: ...Sapphire Infusion Pump User Manual 187 This page is left intentionally blank ...
Page 198: ...Sapphire Infusion Pump User Manual 197 This page is left intentionally blank ...
Page 244: ...Sapphire Infusion Pump User Manual 243 This page is left intentionally blank ...
Page 270: ...Sapphire Infusion Pump User Manual 269 This page is left intentionally blank ...
Page 288: ...Sapphire Infusion Pump User Manual 287 ...
Page 292: ...Sapphire Infusion Pump User Manual 291 This page is left intentionally blank ...
Page 293: ......
Page 294: ......