Sapphire Infusion Pump User Manual
203
Configuring Audio Settings
The
Audio settings
menu allows you to view and modify audio-related pump settings.
Only users with authorization levels of High or Technician have access to this menu.
> To a c c e s s t h e A u d i o s e t t i n g s m e n u :
•
From the toolbar of the
Start Up
screen, select
Options
. Then, select
Pump
configuration
Audio settings
.
Option
Descriptions/Notes
To Modify Parameter
(from the General
settings screen)
Keys volume
Sets the speaker volume for the auditory
signal generated when users select
functions and press keys on the pump.
Select
Keys volume
.
Then, select
Low
,
High
or
Off
.
Alarm volume
Sets the speaker volume for the auditory
alarm signal. Options are
Maximum
or
Minimum
.
When the option is set to
Minimum
,
messages are provided with a visual
signal only. Alarm levels 1, 2 and 3, and
the 'Infusion Complete' message are
provided with a visual signal and the
lowest auditory signal permitted,
according to IEC 60601-1-8.
Maximum alarm volume is 70 dB.
Minimum alarm volume is 56 dB.
Note
: Auditory alarm signal sound
levels, which are less than ambient levels,
can impede operator recognition of
alarm conditions.
For more information about messages
and alarms, refer to
Select
Alarm volume
.
Then, select
Maximum
or
Minimum
.
Summary of Contents for Q Core Sapphire
Page 1: ......
Page 68: ...Sapphire Infusion Pump User Manual 67 This page is left intentionally blank ...
Page 82: ...Sapphire Infusion Pump User Manual 81 This page is left intentionally blank ...
Page 188: ...Sapphire Infusion Pump User Manual 187 This page is left intentionally blank ...
Page 198: ...Sapphire Infusion Pump User Manual 197 This page is left intentionally blank ...
Page 244: ...Sapphire Infusion Pump User Manual 243 This page is left intentionally blank ...
Page 270: ...Sapphire Infusion Pump User Manual 269 This page is left intentionally blank ...
Page 288: ...Sapphire Infusion Pump User Manual 287 ...
Page 292: ...Sapphire Infusion Pump User Manual 291 This page is left intentionally blank ...
Page 293: ......
Page 294: ......