8 Maintenance
Page 16 / 25
8.Maintenance
Autoclavable Components
Heating needle
Autoclavable Procedure
Cleaning
Disinfection
Packing
Sterilization
Storage
Cleaning:
Clean the components with running water with a soft cloth.
Disinfection:
Wipe the components with a piece of gauze that has been
dampened with Ethanol for Disinfection (Ethanol 70 to 80 vol%) and wrung out
thoroughly.
Packing:
Pack the components in
“
Sterilization pouches”.
Sterilization:
Steam sterilization at 134°C at least 4 minutes or
Steam sterilization at 121°C at least 35 minutes.
Minimum drying time after sterilization: 10 minutes.
Storage:
Keep the components in sterilization packaging in a dry and clean
environment.
WARNING
Comply with your national guidelines, standards and requirements for cleaning,
disinfection and sterilization.
Be careful to avoid cross contamination when performing maintenance.
Must be autoclaved after use for each.
Only heating needle can be autoclavable.
NOTE
Heating needle head is easy to damage by external force, take care during
maintenance.