12.EMC Tables
Guidance and manufacturer’s declaration – electromagnetic
emissions
The
CuringPen
is intended for use in the electromagnetic environment
specified below. The customer or the user of the
CuringPen
should assure
that it is used in such an environment
.
Emissions test
Compliance
Electromagnetic
environment-guidance
RF
emissions
CISPR 11
Group 1
The
CuringPen
uses RF
energy only for its internal
function. Therefore, its RF
emissions are very low and
are not likely to cause any
interference
in
nearby
electronic equipment.
RF
emissions
CISPR 11
Class B
The
CuringPen
is suitable
for use in all establishments,
including
domestic
establishments and those
directly connected to the
public low-voltage power
supply network that supplies
buildings used for domestic
purposes.
Harmonic emissions
IEC61000-3-2
Class A
Voltage
fluctuations/flicker
emissions
IEC 61000-3-3
Complies
Guidance and manufacturer’s declaration – electromagnetic
immunity
The
CuringPen
is intended for use in the electromagnetic environment
specified below. The customer or the user of the
CuringPen
should assure
that it is used in such an environment.
Immunity
test
IEC 60601
test level
Compliance
level
Electromagnetic
environment
–
guidance
Electrostatic
discharge
(ESD) IEC 61000-
4-2
+/- 8 kV
contact
+/- 2 kV, +/-
4 kV, +/- 8
kV, +/- 15 kV
air
+/- 8 kV contact
+/- 2 kV, +/- 4 kV,
+/- 8 kV, +/- 15 kV
air
Floors should be
wood, concrete or
ceramic tile. If
floors are covered
with synthetic
material, the
relative humidity