Recommended minimum separation distances
Nowadays, many RF wireless equipments have being used
in various healthcare locations where medical equipment
and/or systems are used. When they are used in close
proximity to medical equipment and/or systems, the medical
equipment and/or systems’ basic safety and essential
performance may be affected. The
Ultra X
has been tested
with the immunity test level in the below table and meet the
related requirements of IEC 60601-1-2:2014. The customer
and/or user should help keep a minimum distance between
RF wireless communications equipments and the
Ultra X
as
recommended below.
Test
Frequency
(MHz)
Band
(MHz)
Service
Modulation
Maximum
power
(W)
Distance
(m)
Immunity test
level
(V/m)
385
380-390
TETRA 400
Pulse
modulation
18Hz
1.8
0.3
27
450
430-470
GMRS 460
FRS 460
FM
± 5 kHz
deviation
1 kHz sine
2
0.3
28
710
704-787
LTE Band
13, 17
Pulse
modulation
217Hz
0.2
0.3
9
745
780
810
800-960
GSM
800/900,
TETRA 800,
iDEN 820,
CDMA 850,
LTE Band 5
Pulse
modulation
18Hz
2
0.3
28
870
930
1720
1700-1990
GSM 1800;
CDMA 1900;
GSM 1900;
DECT;
LTE Band 1,
3,4, 25;
UMTS
Pulse
modulation
217Hz
2
0.3
28
1845
1970
2450
2400-2570
Bluetooth,
WLAN,
802.11b/g/n,
RFID 2450,
Pulse
modulation
217Hz
2
0.3
28