Manual Cleaning Instruction
1. Remove debris
Use a sponge, soft lint-free cloth or soft-bristled brush to assist in removing visible soiling while moving trigger, sleeves,
switches and other moving parts several times. The cleaning brush (Item no. 195011) shown below should be used to
clean handpiece and attachments.
Figure 24
Precautions
• Never immerse the handpiece, lid or attachments in aqueous solutions or in an ultrasonic bath as this could decrease
the service life of the system. The poles in the handpiece must not come into contact with water or solvents: Risk of
short circuit!
• Under no circumstances should the handpiece and the adapter be placed in an immersion bath or in water. Inadvertently
liquid that has penetrated must be removed immediately.
• Do not use pointed objects for cleaning.
2. Disinfection – spray and wipe off
Spray and wipe the device using a neutral pH enzymatic solution for a minimum of 2 minutes.
Follow the enzymatic detergent manufacturer’s directions for correct temperature, water quality (i. e. pH, hardness) and
concentration / dilution.
3. Clean with Detergent
Clean the device manually under running warm water using an enzymatic cleaner or detergent for a minimum of 5 minutes.
Manipulate all moving parts under running water. Use a soft-bristled brush and / or soft lint-free cloth to remove all visible
soil and debris. Follow the enzymatic cleaner or detergent manufacturer’s instructions for use for correct temperature,
water quality and concentration / dilution.
4. After disinfection
Rinse the surface under running water. It is important to ensure that the handpiece and the attachments are at an angle
to prevent water from entering the coupling lugs.
ATTENTION: The cover of the battery housing of the handpiece must be closed so that the poles in the handpiece do not
get caught liquid in contact.
5.
Final Rinse with De-Ionized / Purifi ed Water
Final rinse with de-ionized or purifi ed water for a minimum of 2 minutes.
6. Dry
Dry device using a soft lint-free cloth or clean compressed air.