EIC transfusio-therm 2000 Quick Reference Manual Download Page 5

transfusio-therm

®

 2000      

 

 

Quick Reference Guide 

 

Status: June 2013     Page 5 of 5 

 

20 

Readout of 
the error 
code 
 
 
Error 
message  
 

When switching the appliance on, keep button no. 3 
depressed until the error code appears in the display. Button 
no. 1 lets you browse through the errors all the way to error 
no. 31. To get the number of bags heated, press button no. 2. 
 
Record the read-out error code, together with the serial 
number of the appliance (see nameplate attached on the 
right-hand side of the appliance), in a fault report form and 
fax or e-mail this form to EIC Ltd.  

 

 

EIC Umwelt- und Medizintechnik Ltd., Rudolf-Diesel-Strasse 5, D-37308 Heilbad Heiligenstadt, Germany 

Telephone: +49 3606 607925, Fax: +49 3606 5071353, E-mail: [email protected]

 

Summary of Contents for transfusio-therm 2000

Page 1: ...suggestions when putting the transfusio therm 2000 into operation and when using it I Overview Compartment No 1 Compartment No 3 Compartment No 2 ON OFF Power Switch Start Button with LED LED Alarm Signal Confirmation Buttons with LED Compartment No 1 Compartment No 2 Compartment No 3 Operator Prompting Display Status Line User Interface for Compartment No 1 User Interface for Compartment No 2 Use...

Page 2: ...ion the bag is to be inserted with the connectors located on the right hand side It is possible to thaw a blood bags inserted into protective bags provided the above instructions are observed 04 Closing the door Use the door handle to close the door ON 05 Confirming loading When prompted by the yellow flashing LED and the text appearing in the display press the confirmation button of the compartme...

Page 3: ...mind the operator to resume work on the appliance Whenever an acoustic signal is sounded it is imperative that you go to the appliance and act upon the prompt 14 Safety shutdown If the two channel monitoring system of the appliance detects an error in any compartment area the compartment will be blocked as signaled by a steady red light of the LED ON SERVICE BLOCKED Do not use the blocked compartm...

Page 4: ...with particular care when cleaning temperature measuring points silver plates After cleaning wipe all parts dry and screw the spring plates on again tightly Push the complete rotating unit from the front onto the drive shaft as far as it can go Then reinsert the safety hook Before putting the appliance into use again check that the rotating unit is properly fitted Be sure to clean the motor shaft ...

Page 5: ...rrors all the way to error no 31 To get the number of bags heated press button no 2 Record the read out error code together with the serial number of the appliance see nameplate attached on the right hand side of the appliance in a fault report form and fax or e mail this form to EIC Ltd EIC Umwelt und Medizintechnik Ltd Rudolf Diesel Strasse 5 D 37308 Heilbad Heiligenstadt Germany Telephone 49 36...

Page 6: ...quipment protective cut off Intake 1 10 Temperature difference 1 o C Safety cut off Intake 1 11 Break in the temperature sensor cable Safety cut off Intake 1 12 T1 or T2 39 o C Safety cut off Intake 1 13 Intake leaky HF Safety cut off Intake 1 14 15 16 Full sensor 1 open Full sensor 2 closed Protective cut off Intake 2 17 Full sensor 2 open Full sensor 1 closed Protective cut off Intake 2 18 Tempe...

Reviews: