Ega Master 51266 Series Operating Instructions Manual Download Page 12

12

6. Menu key: long press this key to enter the function setting mode; short press this key again to 

enter next function in order of USB/APO/SEC/CODE/DEF?/ normal measurement; press HOLD key 

to exit menu mode.

7. Data hold key: press this key to lock the measured data display on LCD, and display “HOLD” 

sign at the same time; press this key again to exit hold function.

V SETTING AND OPERATION OF METER

1. Auto power off: power off in about 10 minutes automatically

- Auto power off or not selection setting function: long press “MENU” to enter function selection 

menu, and then short press it to select “APO.0” or “APO.1” interface, and then press “RANGE/+” 

or “MAX/MIN/-” key to switch between “APO.0” and “APO.1”; press “MENU” key to enter next 

function setting or press “HOLD” key to exit function setting, and luxometer will record the 

updated settings.

- Under the state that “APO.0” is selected, auto power off function is closed, namely, the meter 

won’t be power off automatically; under the state that “APO.1” is selected, auto power off function 

is selected, and the auto power off sign is displayed on the screen at the same time. After power off, 

the luxometer will record the settings which will be restored after it is power on, it is unnecessary 

to set up again at each time.

2. Data memory and delete: auto and manual data memory

- The interval of auto data memory is within 0.5-255 seconds, which is adjustable; long press 

“MENU” to enter function selection menu, and then short press it to select SECREC interface, while 

the auto memory interval is displayed on the main display; press “RANGE/+” or “MAX/MIN/-” key 

to adjust the auto memory interval displayed on the main display; press “MENU” key to enter next 

function setting or press “HOLD” to exit function setting. And luxometer will save the updated 

settings.

- Long press “REC/ERASE” to save data automatically, and “REC” sign displayed on LCD SECREC 

flashes, and the luxometer will save the measured data according to the current auto saving interval 

setting; if the storage space of luxometer is fully taken up, the system will exit auto

memory; and “FULL” sign will be displayed on LCD; press “REC/ERASE” key to exit auto data 

memory function.  

- Manual data memory: in the normal measurement state, press “REC/ERASE” once to save one 

measured data, and “REC” sign displayed on LCD SECREC flashes once.

- When 2044 pieces of data is fully recorded, “FULL” sign will be displayed on LCD, and no data 

can be saved at this moment. When the saved data is above 2044 pieces.

- Under power-off state, press “REC/ERASE” and power on meanwhile, when “CLR” sign is 

displayed on LCD, all saved data will be cleared.

3. View saved data:

- Short press “MENU” to enter view recorded data function; if there is no data record in the 

luxometer, “- - - -” sign will be displayed on both LCD main display and secondary display, and 

the luxometer will return to normal measurement state in about 0.5 seconds; If there a certain data 

records in the luxometer, the total number of current data records will be displayed on

LCD secondary display, and the value of the last piece of data will be displayed on the main 

display.

- After entering option of view recorded data, press “UNIT” to view the 1st piece of recorded 

data; press “RANGE/+” and “MAX/MIN/-” keys to view data records forward or backward; press 

Summary of Contents for 51266 Series

Page 1: ...MANUAL DE INSTRUCCIONES OPERATING INSTRUCTIONS LUX METRO LUXOMETER ESPA OL 2 ENGLISH 9 GARANTIA GUARANTEE 17 COD 51266...

Page 2: ...ca la iluminancia recibida en la superficie directa desde la fuente de luz de 1 candela con 1 metro de distancia de la superficie iluminada abreviado como Lux El lux metro puede hacer pruebas de ilumi...

Page 3: ...gas vapor y polvo explosivo No desmonte la cubierta del lux metro sin autorizaci n para evitar da os en el mismo Ponga especial atenci n al sensor de captaci n de luz y mant ngalo limpio y sin rayarlo...

Page 4: ...dad de la lectura entre Lux y FC 3 Bot n de selecci n de valor m ximo o m nimo pulse la tecla para cambiar entre medici n normal m ximo y m nimo El orden de selecci n es medici n normal MAX MIN Cuando...

Page 5: ...ento autom tico que aparece en la pantalla principal Pulse el bot n MENU para cambiar de funci n o pulse HOLD para salir de la funci n de ajustes guardando as los cambios realizados Pulse y mantenga e...

Page 6: ...ngulo y entorno Aseg rese de que no este rayado ni que haya suciedad en el sensor de captaci n de luz de lux metro o los datos de medici n no ser n exactos Cuando la iluminaci n exceda el rango de med...

Page 7: ...alaci n y sustituci n de la bater a Cuando el indicador de la bater a aparezca en la pantalla del lux metro reemplace la bater a inmediatamente Sustituya la bater a de la siguiente manera Apague el lu...

Page 8: ...deben ser recogidos y dispuestos por separado Si usted tiene que tirar por favor no use la basura habitual Por favor p ngase en contacto con su distribuidor para el reciclaje de forma gratuita GARANT...

Page 9: ...ed as Lux Cod 51266 series can test illuminance within 0 20 000Lux II UNPACKING INSPECTION Unpack and check if there is any damage or loss contact with the nearest local sales service agency Mainframe...

Page 10: ...losive gas steam and dust Do not dismantle the shell of luxometer without authorization to avoid luxometer damage Do not charge the battery to avoid battery explosion and personal injury Please pay at...

Page 11: ...ing between Lux and FC 3 Maximum value and minimum value selection key under measuring state press the key to switch between normal measurement MAX and MIN in order of normal measurement MAX MIN When...

Page 12: ...splayed on the main display press RANGE or MAX MIN key to adjust the auto memory interval displayed on the main display press MENU key to enter next function setting or press HOLD to exit function set...

Page 13: ...ting source and illuminometer positions or different of them or the environment is different then the measured data will be different cause the intensity of light varies with change of distance positi...

Page 14: ...20FC 0 01FC 200FC 0 1FC 2000FC 1FC VII MAINTENANCE Warning During replacing battery or measuring keep the sensitive mirror from scratch or dirt to avoid damaging the mirror or influencing the measuri...

Page 15: ...ibility for damage or malfunction as a result of the device being incorrectly used or applied for a purpose for whith it was not intended According to Waste Electrical and Electronic Equipment directi...

Page 16: ......

Page 17: ......

Page 18: ......

Page 19: ......

Page 20: ......

Reviews: