TROUBLE-SHOOTING THE ADC-T

(1) Verify that your USB com driver is installed by going to Device Manager and checking for the Prolific USB to
Serial Comm Port entry. Right click on the entry, then Properties, select the Driver tab and verify that the driver is
dated 8/15/2014 or later. If the driver is out of date, connect to the Internet and click the Update Driver button. Please
note that an out of date driver may allow the ADC-T to partially function and/or with erratic operation.

(2) Verify that the Prolific USB to Serial Comm Port entry is the com port used by your ADC-T. You may do this by
watching the entry in Device Manager and unplugging your ADC-T from your computer USB port. The entry should
disappear and then re-appear when you plug the ADC-T back in to your computer USB port.

(3) Try replacing the USB cable and/or using a different USB port.

(4) Verify that the ADC-T device driver is correctly installed and running. Click the blue icon in your system
tray (bottom right) and check for the correct com port setting (should show the correct com port open). If the window
below the COM indicator shows "COM Not Found!" then a com port issue is the problem (check for another program
that may be using the com port). You may right click the taskbar at the bottom of your screen, then Task Manager to
view running applications.
.
(5) Remove the ADC-T USB cable from your computer, wait 5 seconds and plug back in. Open the ADC-T
User Interface by clicking on the blue icon in your system tray and double click on the blue EECI logo to reset
the com port. You should see a reset message followed by a COM open indication. Click Setup and uncheck "Allow
External Commands". If this corrects the issue that you are experiencing then an external application is the problem.

(6) If erratic operation is experienced, check for loose connections at the terminal block (tug on each wire going into 
the terminal block), check for short circuits caused by metal contact to the ADC circuit areas or other connected 
hardware.

(7) The input voltage to any of the analog inputs must not exceed the reference voltage (nor should the input go
negative with respect to the reference when not in differential mode) or erratic data may be generated on one or more 
of the analog channels. When an external reference is used, the reference voltage must not exceed 5 volts. 

(8) When a temperature input is left unconnected, the ADC-T temperature Interface will display a blank indication for 
that temperature input. When a temperature input is shorted, a very low negative temperature will be displayed.

Please contact EECI Support at (800) 842-7714 or (937) 349-6000 if you require additional assistance or have
questions.

 

EXPANSION PORTS

Page 5

The ADC-T provides three expansion ports for additional I/O operations using the same COM port. The
expansion ports may used to connect up to 144 relays (8 per expansion port) or to connect up to 144 inputs (or
combinations of relays and inputs) using EX-8M and/or EXM-16 expansion modules. The relay output ports on the
EX-8M and EXM-16 are identical to those used on all EECI products and accept the same relay cards and other 

accessories. A variety of relay cards with various contact configurations are available for connection to the expansion 
modules. A partial list of relay cards includes the RD-8, RD-8M, RG-8, RH-8, RI-8, RJ-8, RL-8, RN-8, RN-8M, RP-8 
and RJ-8. The RYD-8 relay driver card may be used to control larger power relays, contactors, solenoids, valves or 
other devices. A Digital to Analog expansion module may be used to output a 0 to 5 volt or user supplied analog 
level. For additional information on the EX-8M Expansion card, please refer the the EX-8M Technical Reference 
available on the EECI web site.

(Step 5) After you have verified

 or installed the up to date USB Com driver, reconnect (if not already connected*) your 

ADC-T Analog to Digital Converter to the same USB port and install the ADC-T device driver** by clicking the "Install 
EECI Device Driver" button from the installer dialog window (if you have auto start disabled, you may browse to the 
Windows Driver folder and double click on setup.exe). A desktop icon for the ADC-T will be created. When installing the 
ADC-T device driver on a Windows XP or Vista computer, double click on setup.exe in the XP-Vista folder.             
*if you reconnect your ADC-T to a different USB port, a different com port may be assigned.

**Please note that the Microsoft .Net Framework must be installed on your system. The .Net Framework is already 
installed if you have Windows 7 or higher. If the .Net Framework is not installed on your computer, the installation 
program will attempt to download and install the .Net Framework from the Microsoft Update website through your 
internet connection. This may take 5 minutes or longer. In order for the .Net Framework to install correctly, your 
computer must be up to date with Windows Update. This is especially important with Windows XP* and Vista. You may 
update your computer by clicking on the start button, All Programs, Windows Update and clicking on "Check for 
updates". *for XP, Service Pack 3 must be installed.

(Step 6) After the ADC-T device driver is installed

, the 

ADC-T

 com port should be detected and a pop-up will prompt 

you for the com port to be used with your 

ADC-T

 A to D Converter (correct com port should 

be shown).You may edit the 

com port if incorrect. The ADC-T User Interface will now open and the label just under the COM port indicator should 
show that Com port open. When the sensors are left unconnected your  screen should show a row of 4, 8 or 12 blank 
boxes. Click the "Install Help Files" button to install important trouble-shooting and reference information before you 
close the installer.

you may label your temperature inputs and

The ADC-T Data Logger may be started by checking the Enable box. The data log entry interval may be set to minutes 
or seconds (click the M or S button) and entering the desired time interval. The settings will not go into effect until the 
Apply button is clicked. 

Click the "Tray" button to place the User Interface in "driver 
only" mode. You

 

will see a blue 4 icon in your system tray by 

the clock (bottom right of screen as shown below). 

Click the blue 4 icon to re-open the ADC-T User 

Interface. Click  "Unload Device Driver" to unload the ADC-T device driver.

Please note that by default, the ADC-T device driver is only loaded on demand to conserve system resources. To re-
start the ADC-T device driver, double click the red and blue ADC-T desktop icon or under Applications.

(Step 7) Test each temperature input

 by connecting a sensor (one at a time) to each of the +temperature inputs and a 

GND (-)  terminal. Use caution to observe the correct sensor polarity or damage could result. Your screen should show 
room temperature for that input. When you press the sensor between your forefinger and thumb, you should see the 
temperature increase. Please see page 8 for the temperature sensor TO-92 package pin connections.

If operation of the temperature inputs are normal, 

 calibrate as described on 

page 10.Testing is now complete and your ADC-

T

 may be placed in service. If problems are encountered during testing, 

proceed to the trouble-shooting procedures shown on page 5.

HOW TO OPEN DEVICE MANAGER

Windows XP, VISTA or Windows 7

Click the Start Button (lower left of screen), then Control Panel (right side). With the view set to classic view, small or
large icons, click (or double click) the Device Manager icon. With Windows XP you will need to click the System icon
(in Control Panel) then the Hardware Tab then the Device Manager button. You may also use the supplied CD to open
Device Manager by clicking the "Open Device Manager" button when the CD auto starts.

Windows 8.1, 10 and Windows Server

Move your mouse cursor to the lower right side of your screen and click on Settings. Click on Control Panel near the
top and click on the Device Manager icon.

Page 4

Summary of Contents for ADC-T Series

Page 1: ...d range sensors 67F to 302F 55C to 150C Resolution 1 degree optional 1 4 degree resolution available Accuracy 1 degree when using the standard sensors Temperature Inputs Available in 4 8 and 12 channe...

Page 2: ...nager The entry should disappear and then reappear when you plug the ADC T back in If you do not have an Internet connection or are having problems with the USB Com driver installation disconnect the...

Page 3: ...ADC T Analog to Digital Converter to the same USB port and install the ADC T device driver by clicking the Install EECI Device Driver button from the installer dialog window if you have auto start dis...

Page 4: ...t shown The com port may be changed by clicking the Setup button and changing the com number in the Com Port window and clicking Apply Clicking the Setup button allows you to label the temperature inp...

Page 5: ...ion 4 5 volt allows a 0 to 5 volt input range or 2 5 to 2 5 volts in differential mode with reference 3 channel single ended reference voltage from AUX header is applied to the channel 4 analog input...

Page 6: ...f 0 volts at 0 degree K This is how the offset of 459 degrees is obtained since degrees K is 459 less then degrees F or 273 degrees less then degrees C The scale is determined by dividing the 4 096 vo...

Page 7: ...idge Rd Milford Center OH 43045 9016 USA 937 349 6000 800 842 7714 937 349 6000 sales eeci com https www eeci com PHONE ORDERS TECH SUPPORT E mail Web International Domestic Copyright 2014 2021 Electr...

Reviews: