EECI ADC-4U15 Technical Reference Download Page 7

RAW ANALOG DATA

The ADC-4U15 will present the raw analog data as a 
number between 0 and 32767* when in 15 bit single 
ended mode and a number between -32768 and 
+32767* when in 16 bit differential mode.

*the usable range with the ADC-4U15 reference
voltage set to 6.144 will be 0 to 26665 in 15 bit single 
ended mode and -26666 to +26665 in differential mode
since the portion of the ranges from (5 volts to 6.144) 
and from (-6.144 to -5 volts) are not usable.

HIGH PRECISION

 

INSTRUMENT AMP

The built-in high precision instrument amplifier utilizes
6 predefined precision voltage references that have 
been selected to provide an exact voltage 
measurement per increment as follows:

256 mv

.0078125 mv increments

(128 per mv)

512 mv

.015625 mv increments 

(64 per mv)

1.024 v

.03125 mv increments 

(32 per mv)

2.048 v

.0625 mv increments 

(16 per mv)

4.096 v

.125 mv increments 

(8 per mv)

6.144 v

.1875 mv increments 

(16 per 3 mv)

DIFFERENTIAL INPUT IMPEDANCE

at 6.144 volts:

22 Meg ohm

at 4.096 volts:

15 Meg ohm

at 2.048 volts:

4.9 Meg ohm

at 1.024 volts:

2.4 Meg ohm

     at 512 mv:

710 Kilo ohm

     at 256 mv:

710 Kilo ohm

NOTE: 

Once a change is made, the Apply Button will 

appear. Changes will not be applied until you click the 
Apply button. The changes are saved to the 
ADC-4U15.ini file.

Relay and Input labels may be changed by clicking the Relay Control or Input Setup buttons. Click the Analog Setup 
button to make changes to the analog labels, suffix, scale, offset or precision. These also apply to logged data.

The voltage input range may be set by selecting 1 of the 6 possible gain levels using the built in instrument amp.

Start the data logger by checking the Enable box. The logger will roll-over or stop based on the setting in the Max 
Entries box. The Data Interval may be set to minutes or seconds and will log the data based on the setting in the box.

Click the Com Port button to refresh the available Com Ports. Enter or use the pull-down arrow to set the Com Port.

The Sample Rate is the interval between samples (in the example above, the setting is 100 milliseconds or 10 
samples per second).

Check the Start Minimized box to start the ADC-4U15 device driver in "driver only" mode. Blue 4 icon will show in 
your system tray.

Page 12

ELECTRONIC ENERGY CONTROL, INC.

14960 Maple Ridge Rd

Milford Center OH 43045-9016

USA

(937) 349-6000
(800) 842-7714
(937) 349-6000
[email protected]
http://www.eeci.com

PHONE*.................
ORDERS................
TECH SUPPORT...
E-mail*....................
Web*........................
            *International & Domestic

Copyright      2014 - 2022  Electronic Energy Control, Inc.  All Rights Reserved

c

®

Summary of Contents for ADC-4U15

Page 1: ...it resolution no missing codes 15 bit resolution will divide your measurement range into 32 768 different increments 16 bit resolution will divide your measurement range into 65 536 different incremen...

Page 2: ...USB port and install the ADC 4U15 device driver by clicking the Install EECI Device Driver button from the installer dialog window if you have auto start disabled you may browse to the Windows Driver...

Page 3: ...sion modules The relay output ports on the Page 5 installed if you have Windows 7 or higher If the Net Framework is not installed on your computer the installation program will attempt to download and...

Page 4: ...alog converter may be used to output an analog voltage For additional information please visit t the EECI web site at www eeci com THE ADC 4U15 USER INTERFACE The ADC 4U15 User Interface allows you to...

Page 5: ...A 8 instrumentation amplifier The VA 1 connects to any of the analog inputs and will convert a millivolt signal input to a 0 to 5 volt output for input into the ADC Signal levels this low will may req...

Page 6: ...24 volts or great enough to create an electrical shock hazard or fire should be connected to the ADC analog input using a step down transformer to reduce the high voltage down to the 5 volt range as...

Page 7: ...es will not be applied until you click the Apply button The changes are saved to the ADC 4U15 ini file Relay and Input labels may be changed by clicking the Relay Control or Input Setup buttons Click...

Reviews: