Edwards VolumeView  EV1000 Setup Manual Download Page 2

EV1000 Clinical Platform Startup with the VolumeView Set

Before startup, please ensure monitor, cables and patient cables

are all connected.

1.  Turn on monitor and databox. System self-test starts.
2.  Enter new patient data OR continue with existing patient.
3.  Set up and prime VolumeView sensor and TruWave CVP

  transducer and connect to databox.
4.  Ensure the sensor and transducer

  are level to patient’s phlebostatic

  axis and zero transducers on monitor

 using 

Clinical Actions button.

5.  Choose continuous parameters,

  and defi ne alarms/target range.
6.  Start transpulmonary thermodilution - 

inject bolus and review and edit boluses 

results.

7.  Monitor patient in real time with one of the six available

 screens.
Thermodilution Bolus - VolumeView Set
1.  Insert the VolumeView femoral arterial catheter

  according to the directions for use.
2.  Insert a Central Venous Catheter (CVC) (if using the Edwards 

oxymetry CVC , refer to the Edwards oxymetry CVC setup 

guide).

3.  Ensure the TruWave central venous pressure

  transducer and VolumeView sensor are primed

  and then connected to the EV1000 databox.
4.  Ensure the sensor and transducer

  are level to patient’s phlebostatic

  axis and zero transducers on monitor

  using Clinical Actions button.
5.  Connect the VolumeView sensor

  to the VolumeView femoral arterial

  catheter and confi rm the appropriate

  waveforms and pressures are displayed.

6.   Connect the databox thermistor cable to the VolumeView

 

femoral arterial thermistor and confi rm appropriate 

 

temperature on the EV1000 monitor.

7.   Prime the VolumeView set manifold, removing all air.

 

Connect the TruWave central venous pressure transducer

 

to the VolumeView manifold and then connect it to 

 

the central venous catheter and 

 

confi rm the appropriate waveform 

and pressure are displayed.

8.   Connect the in-line injectate 

 

thermistor probe to the VolumeView 

manifold.

9.   Connect a pre-chilled syringe

 

with (15 or 20 ml) normal saline to

 

the VolumeView manifold valve/port.

10.  Touch the 

Clinical Actions button 

and then touch the 

Thermodilution 

 

button to start the TPTD procedure.

11.  Select the desired injectate volume on

 

the keypad (prefer 15 or 20 ml).

12.  Select 

Lung Resection, if appropriate,

 

as well as the descriptor (for example

 

RUL = right upper lobectomy).

13.  Select 

Parameters, to choose indexed or 

non-indexed parameter.

14.  Touch 

Start Set.

15.  

Wait is highlighted until the thermal baseline is reached.

16.  When 

Inject is highlighted, use a rapid, smooth, continuous

 

method to inject the cold solution with the volume amount

 

previously selected.

17.  Observe the calculation.
18.  Repeat steps 14 and 15.
19.  Touch review to edit the series of boluses.
20.  Touch 

Accept.

8