STP-iXA3306 Series Turbomolecular Pump
MT-82E-001-H
Page 60
OPER
A
T
ION
4.6.2
Starting/stopping the STP pump
There are two methods for the starting/stopping operation with the parallel port.
Use one of them.
1. Contact input
Open the circuit between (19)
–(37).
Method
Starting the pump
Stopping the pump
1
1) Short the circuit between (1)-(21).
2) Short the circuit between (3)-(21) for 0.3
seconds or more.
However, when inputting this signal
simultaneously with switching "ON" the STP
pump, short the circuit between (3)-(21) for 10
seconds or more.
Open the circuit between (1)-(21).
2
Short the circuit between (1)-(3).
In this case, (21) is not used.
Open the circuit between (1)-(3).
Table 14 - Starting/stopping the STP pump using contact input (X2 REMOTE)
0.3 sec. or more
Stop
Rotation
Stop
停止
CLOSE
OPEN
(1)
–(21)
(3)
–(21)
X2 REMOTE
connector
Pump operation
OPEN
CLOSE
Figure 31 - Pump operation method 1 (contact input)
Stop
Rotation
Stop
(1)
–(3)
X2 REMOTE
connector
OPEN
CLOSE
Pump operation
Figure 32 - Pump operation method 2 (contact input)