6.4 System operating temperature configuration
The system operating temperature configuration can be adjusted to suit process
demands. Contact your supplier for instructions on how to access the configuration
menu (through the PDT) and for process specific recommendations.
Use the following procedure:
1.
Access the Config menu on the PDT (requires an access code, contact supplier).
2.
Scroll to Temp Set-points - press Enter.
3.
Scroll to Set DP Temp - press Enter.
4.
Scroll to desired temperature - Press Enter.
Note:
The Dry Pump set-point is the desired surface temperature of the internal low
vacuum stage.
Refer to
Table: Dry pump and booster temperature settings
for default temperature
settings and ranges.
Table 32
Dry pump and booster temperature settings
Temperature Setting
Range in which temperature can be
set(°C)
Default value (°C)
Dry pump
(Standard variants)
iXM600, iXM1200
60 - 110
80
Dry pump
(Standard variants)
iXM200, iXM3000
60 - 110
110
Dry pump
(T and TX variants)
60 - 130
130
Booster
(Only TX variants)
-
85
10/2020 - ©Edwards Limited
Page 65
M56635880_D
M56635880_D - Operation
Summary of Contents for iXM Series
Page 1: ...Dry Pump System iXM INSTRUCTION MANUAL M56635880_D Original instructions edwardsvacuum com...
Page 101: ...This page has been intentionally left blank...
Page 102: ...This page has been intentionally left blank...
Page 103: ...This page has been intentionally left blank...
Page 104: ...edwardsvacuum com...