Technical data
A541-11-880 Issue A
2.3
Electrical data
Table 4 - General electrical data
Description
Rating
Units
Input supply voltage, 3-phase, a.c., rms
200 - 460 ±10%
V
Input supply frequency
50 - 60 ±5 Hz
Hz
Input supply wiring configuration
3 wire plus Earth (ground)
Input supply branch circuit protection requirements
Refer to
Input supply voltage un-balance
Voltage un-balance should NOT exceed 2%, when
assessed over any one minute period.
%
Input supply short-circuit rating
30 (absolute maximum rating)
kA
System Earthing
200 - 240 V: TT; TN and IT
(including corner-Earth supplies)
380 - 460 V: TT; TN and IT
(excluding corner-Earth supplies)
Second protective Earth (ground) conductor
Must be fitted with a cross-sectional area at least
equal to the phase conductor size.
Installation category
Over-voltage category II (IEC 60664-1)
Typical Earth leakage
*
*
Typical Earth leakage values were measured at steady-state conditions.
Note that higher leakage currents may occur:
i) under transient conditions such as power on or pump acceleration or
ii) with certain supply configurations such as Earthed phase or
iii) with abnormal supply configurations such as a missing phase or unbalanced supply voltages
Contact Edwards for more information about configuration requirements for Earth leakage reduction.
<3.5
mA
Table 5 - Electrical ratings: ultimate vacuum and extracted (700 torr) exhaust load condition
Electrical sub-system: ultimate vacuum (0.01 mbar) load rating
*
*
The electrical sub-system ratings, detailed in
, assume an input short-circuit rating of 30 kA.
Input supply
voltage
Input supply
frequency
Input supply current
Input supply
power
Displayed power / motor
power
†
†
The displayed power consumption, in
, is a typical value and does NOT include the power
consumption of the pump’s internal auxiliary power-supply. This auxiliary power-supply is used to power the
pump’s motor-controller, pump-controller and customer accessories.
(V) (a.c., rms)
(Hz)
(A) (a.c., rms)
(A)
(Repetitive
Peak)
(kW)
(kW)
200
50
4.0
16.2
0.55
0.52
200
60
3.9
15.2
0.55
0.52
240
50
3.3
13.6
0.55
0.52
240
60
3.2
12.9
0.55
0.52
380
50
2.2
9.0
0.55
0.52
380
60
2.1
8.8
0.55
0.52
460
50
1.8
7.7
0.55
0.52
460
60
1.8
7.5
0.55
0.52