Installation

3

Installation

3.1

Unpack and inspect

Remove all packing materials and check the FL20K. If the FL20K is damaged, notify your supplier and carrier in writing 
within three days; state the item number of the FL20K together with your order number and your supplier’s invoice 
number. Retain all packaging materials for inspection. Do not use the FL20K if it is damaged.

Check that your package contains the items shown in 

Table 1

. If any one of these items is missing, notify your supplier 

in writing within three days.

If the FL20K is not to be used immediately, return it to it’s protective packaging.

3.2

Install the FL20K

We recommend that you mount the FL20K horizontally, as shown in 

Figure 1

. If necessary, however, you can mount 

the FL20K in any orientation, as the sorbent is securely retained in the basket. The mass of the FL20K will be 
adequately supported by the pipeline connections and you can connect either port to the rotary vacuum pump. 
However, we recommend that you mount the FL20K as close to the vacuum system and as far from the rotary pump 
as possible, to maximise the useful life of the sorbent. When you mount the FL20K, ensure that you leave adequate 
clearance to enable the sorbent basket to be removed.

Use the standard KF coupling components (that is, NW clamping rings, centring rings and O-rings) to connect the 
FL20K to the pump and the vacuum system.

3.3

Fill the FL20K with sorbent

CAUTION

Do not overfill the basket with sorbent, If you do, crushed sorbent could be sucked into the rotary pump and 
damage the pump.

1. Refer to 

Figure 1

; unscrew and remove the nut (1) and the lid (6).

2. Remove the basket (4) and spring (5) from the trap body (3).

3. Refer to 

Figure 3

. Remove the wing nut (9) and end cap (1) from the basket (3). Remove the washer (10), 

compression spring (8) and perforated disk(7).

4. Fill the basket with sorbant; when full, the end disk (7) should align with the marker (6) on the spindle:

Table 1 - Checklist of items

Qty

Description

Check (

)

1

FL20K  Valve

1

Tin of activated alumina

WARNING

Activated alumina is a desiccant and will absorb materials from your process system. You must take 
suitable precautions to protect people from the harmful effects of any noxious materials released 
if you bake the spent alumina to reactivate it. Avoid inhalation of its dust, eye contact and 
unnecessary skin contact.

Summary of Contents for FL20K

Page 1: ...A133 99 880 Issue E Original Instruction Manual FL20K Foreline Trap Description Item Number FL20K A133 05 000...

Page 2: ...This page has been intentionally left blank...

Page 3: ...cal data 3 3 Installation 4 3 1 Unpack and inspect 4 3 2 Install the FL20K 4 3 3 Fill the FL20K with sorbent 4 4 Operation 6 5 Maintenance 7 5 1 Replace the sorbent charge 7 6 Storage and disposal 8 6...

Page 4: ...13 All rights reserved Edwards and the Edwards logo are trademarks of Edwards Limited Contents Tables Table Page 1 Checklist of items 4 Associated publications Publication title Publication number Vac...

Page 5: ...oughout this manual conform to the SI international system of units of measurement 1 2 General Description The FL20K has been developed for use in clean pumping systems to prevent the back migration o...

Page 6: ...All rights reserved Edwards and the Edwards logo are trademarks of Edwards Limited Introduction Figure 1 Component parts of the FL20K 1 Nut 2 Wing nut 3 Body 4 Basket 5 Compression spring 6 Lid 7 O ri...

Page 7: ...ue E 2 Technical Data 2 1 Operating conditions 2 2 Performance 2 3 Mechanical data Figure 2 Dimensions mm Ambient operating temperature range 0 to 50 C Leak tightness 1 bar 1 x 105 Pa differential 1 x...

Page 8: ...e FL20K as close to the vacuum system and as far from the rotary pump as possible to maximise the useful life of the sorbent When you mount the FL20K ensure that you leave adequate clearance to enable...

Page 9: ...rests on the end disk 7 Ensure that both end disks 5 and 7 have their smooth side in contact with the balls 6 Refit the washer 10 and end cap 1 over the spindle 7 Refit the second washer 10 and wing...

Page 10: ...d pump down If the FL20K has been open to atmosphere for a long period of time it may be necessary to carry out prolonged pumping with a gas ballasted rotary pump to remove the absorbed moisture Alter...

Page 11: ...every three months as a routine maintenance procedure 1 Refer to Figure 1 Unscrew and remove the nut 1 and the lid 6 2 Remove the basket 4 and spring 5 from the trap body 3 3 Refer to Figure 3 Remove...

Page 12: ...Limited Storage and disposal 6 Storage and disposal 6 1 Storage 1 Return the FL20K to its protective packaging 2 Store the FL20K in clean dry conditions away from direct sources of heat and sunlight 6...

Page 13: ...tributor When you order please state for each part required Model and Item Number of your equipment Serial Number if any Item Number and description of part 7 2 Spares 7 3 Accessories Description Item...

Page 14: ...A133 99 880 Issue E Page 10 Edwards Limited 2013 All rights reserved Edwards and the Edwards logo are trademarks of Edwards Limited This page has been intentionally left blank...

Reviews: