Technica
l data
D396-14-880 Issue J
2
Technical data
2.1
Operating and storage conditions
2.2
Mechanical data
2.3
Electrical data
Table 1 - Operating and storage conditions
Ambient operating temperature range
0 to 35 °C
Ambient storage temperature range
-20 to 40 °C
Maximum ambient operating humidity
10 to 95% RH (non-condensing to DIN 40040)
Maximum operating altitude
3000 m
Table 2 - Mechanical data
Dimensions
Mass
Connectors
See
Table 3 - Logic interface electrical data
Relay contact rating (resistive load)
1 A at 25 V d.c.
Remote control signals
Control voltage: low (close)
< 0.8 V d.c.
Control voltage: high (open)
4 to 24 V d.c.
Maximum input current (at 24 V)
80
μ
A
Maximum output current (at 0 V d.c.)
160
μ
A
Air-cooler electrical supply
Vent-valve electrical supply
Analogue EXT pump speed output
Output voltage
0 to +5 V d.c., proportional to EXT speed
Maximum output current
5 mA
Table 4 - Fuse ratings
Electrical supply fuse
EXC120/EXC120E
3.15 A, type T 20 mm
EXC300/EXC300M
6.3 A, type T 20 mm
Bakeout band fuse
*
*
Not applicable to the EXC120E Controller.
90 to 132 V a.c. electrical supply
1 A, type F 20 mm
180 to 264 V a.c. electrical supply
0.5 A, type F 20 mm