Installation
3.10
Adjust the Normal speed
Note:
If you set the Normal speed to be more than 70% of full rotational speed, the TMP Normal output will go
high when you select Standby. The Controller is supplied with the Normal speed set to 80% of full rotational
speed.
You can adjust the Normal speed (at which the TMP Normal output goes low: see
) between 65 and 95%
of full rotational speed.
. To adjust the Normal speed, use a small screwdriver to turn the SETPOINT potentiometer (A).
also shows the approximate Normal speed settings for different potentiometer adjustments.
3.11
Adjust the Internal Timer
The Internal Timer can be adjusted between approximately 1 and 30 minutes. The Controller is supplied with the
Internal Timer adjusted to eight minutes (see
).
To adjust the Internal Timer, use a small screwdriver to turn the TIMER potentiometer (B).
also shows the
approximate timer settings for different potentiometer adjustments; these settings are only approximate. If the
timer setting is critical for your application, you must check the time and readjust the potentiometer as necessary,
until the correct time is obtained.