A652–07–880 Issue D
Page 10
© Edwards Limited 2014. All rights reserved.
Edwards and the Edwards logo are trade marks of Edwards Limited.
Installation
3.3
Unpack and inspect
Remove all packing materials and protective covers and inspect the equipment. If any of the equipment is damaged,
notify the supplier and the carrier in writing within three days. State the Item Number of the equipment together
with the order number and the supplier’s invoice number. Retain all the packing materials for inspection. Do not use
the equipment if it is damaged.
Check that the items listed in
have been received. If any item is missing, notify the supplier in writing within
three days.
If the E-LAB 2 is not to be used immediately, replace the protective covers. Store the E-LAB 2 in suitable conditions,
as described in
.
3.4
Locate the pump
The E-LAB 2 has a lifting handle that can be used to move the pump by hand. If using mechanical lifting equipment,
do not attach the equipment to the handle; for stability, use slings around the motor and the pump body.
Provide a firm, level platform for the E-LAB 2. Locate the E-LAB 2 so that the oil level sight glass (
, item 8)
and the oil change tag (10) are visible, and so that the oil filler plug (7) and oil drain plug are (9) accessible.
If the E-LAB 2 will be located inside an enclosure, make sure that there is adequate ventilation at both ends of the
pump, so that the ambient temperature around the pump does not exceed 40 °C. There must be a minimum space
of 25 mm between the pump and the enclosure walls.
Table 3 - Checklist of components
Qty
Description
Check (
)
1
E-LAB 2 Pump (with oil change tag fitted)
1
Oil mist filter
1
Inlet catchpot
6
Clamping rings
4
Co-Seals
1
Centring ring and O-ring
2
Elbows
2
Hose adaptors
1
Electrical supply cable *
Gas ballast oil return kit, which contains:
1
Oil return assembly
1
Drain adaptor
1
Restrictor
1
Flexible oil-return tube
2
Hose clips
*
Not supplied with pumps with Item Number A652-07-903; see
.