TECHNICAL DATA
D141-32-880 Issue A
2.2
Performance
Gases detected:
The GasCheck G3 can detect any gas or vapour which has a different thermal conductivity to the ambient air in which
it was zeroed. The larger the difference the greater the sensitivity, so that GasCheck G3 is not highly sensitive to the
gases normally found in large concentrations in the ambient air, e.g. N2 (nitrogen) and O2 (oxygen).
Note:
The GasCheck G3 cannot identify different gases and it cannot differentiate between different gases.
The GasCheck G3 comes supplied with an extensive list of calibration factors for common gases (refer to
). In
addition, the GasCheck G3 comes with 5 generic 'gas groups' with a variation in sensitivities which can be used as a
substitute for gases not listed in
(group 5 being the highest sensitivity through to group 1 being a low
sensitivity); Contact Edwards for advice on which groups to use.
Response time:
Detection time to T90
1 sec (short probe), 9 sec (long probe)
Clear down time
1 sec (short probe), 9 sec (long probe)
Minimum detectable leak
Refer to
Table 1 - Smallest detectable leak levels
Name
Abbreviation
Minimum sensitivity
cc/sec
Hydrogen
H2
7.7 E-6 cc/sec
Helium
He
1.0 E-5 cc/sec
Refrigerant R12
R12
2.7 E-5 cc/sec
Refrigerant R1301
R1301
2.4 E-5 cc/sec
Refrigerant R134a
R134a
5.8 E-5 cc/sec
Refrigerant R22
R22
2.6 E-5 cc/sec
Refrigerant R11
R11
3.2 E-5 cc/sec
Sulphur Hexaflouride
SF6
2.2 E-5 cc/sec
Carbon dioxide
CO2
4.0 E-5 cc/sec
Methane
CH4
2.9 E-5 cc/sec
Argon
Ar
3.5 E-5 cc/sec
Oxygen
O2
2.9 E-4 cc/sec
Refrigerant R502
R502
3.0 E-5 cc/sec
Refrigerant R404a
R404a
3.2 E-5 cc/sec
Refrigerant R407c
R407c
3.3 E-5 cc/sec
Refrigerant R410a
R410a
3.2 E-5 cc/sec
Refrigerant R507
R507
3.8 E-5 cc/sec