© Edwards Limited 2009. All rights reserved.
Page 13
Edwards and the Edwards logo are trademarks of Edwards Limited.
OPERATION
D141-32-880 Issue A
Figure 9 - Change between short and long probes
. Slide the box spanner over the short probe (A) as shown in Figure 9 (B). Unscrew the probe counter-
clockwise. Remove the spanner and the probe assembly (C). To replace the short probe with the long probe, perform
the same sequence in reverse order.
Note:
When detecting leaks with the long probe fitted, the response time of the GasCheck G3 is significantly
increased. The speed at which the probe is drawn over the inspection surface should be reduced to around
10mm per second.