Page 10
W601100880_A - Technical data
Electrical connection
FCC68/RJ45, 8 pin socket or D-Sub, 9-pin,
male
Sensor cable
8-pin or 9-pin plus shielding
Cable length
§
100 m (0.14 mm² conductor)
For longer cables, larger conductor cross-sections are required.
Grounding concept
Vacuum flange - signal common
See
Supply common - signal common
Conducted separately; for differential
measurement (10
)
*
The gauge is protected against reverse polarity of the supply voltage.
§
For longer cables, larger conductor cross sections are required.
Table 5
Materials exposed to vacuum
Materials exposed to vacuum
Flange, tube
Stainless steel AISI 316L
Sensor, feedthrough
Ceramics (Al
2
O
3
99.5%)
Internal volume
DN 16 ISO-KF
3.7 cm
3
8 VCR female
5.1 cm
3
Admissible pressure (absolute)
00 Torr/mbar (F.S.)
4 bar
100 Torr/mbar (F.S.)
3 bar
10 Torr/mbar (F.S.)
2 bar
Bursting pressure (absolute)
6 bar
Table 6
Environmental
Temperature compensated range
+10 °C to +50 °C
Admissible temperatures
Storage
-20 °C to +85 °C
Operation
0 °C to +70 °C
Bakeout (not in operation)
110 °C at the flange
Relative humidity
80% at temperatures
+31 °C decreasing to 50% at +40 °C
Use
Indoors only, altitude up to 4000 metres NN
Degree of protection
IP 40
Table 4
Electrical (continued)
Summary of Contents for Barocel 7025M
Page 1: ...W601100880_A Original Instructions Barocel 7025M Gauge INSTRUCTION MANUAL edwardsvacuum com ...
Page 6: ...This page has been intentionally left blank Page 4 W601100880_A Contents ...
Page 13: ...Page 11 W601100880_A Technical data Dimensions Figure 1 Dimensions mm ...
Page 25: ...This page has been intentionally left blank ...
Page 26: ...This page has been intentionally left blank ...
Page 27: ......
Page 28: ...edwardsvacuum com ...