A726-01-840
Issue F Original
Instruction Manual
XDS Replacement Tip Seal and Exhaust Valve Kit
Description
Item Number
XDS5/10 Replacement Tip Seal Kit
A726-01-805
Page 1: ...A726 01 840 Issue F Original Instruction Manual XDS Replacement Tip Seal and Exhaust Valve Kit Description Item Number XDS5 10 Replacement Tip Seal Kit A726 01 805...
Page 2: ......
Page 3: ...t valve 1 For return of equipment complete the HS Forms at the end of this manual Illustrations Figure Page 1 Replacing the tip seals pre serial number 037874870 XDS5 or 037869927 XDS10 3 2 Replacing...
Page 4: ...This page has been intentionally left blank A726 01 840 Issue F Page ii Edwards Limited 2010 All rights reserved Edwards and the Edwards logo are trademarks of Edwards Limited...
Page 5: ...XDS pumps post serial number 037874870 XDS5 or 037869927 XDS10 continue at Step 9 4 The exhaust duct 105 is now exposed on the outside of the fixed scroll If the pump is an XDS C remove the 5 screws 1...
Page 6: ...centre of the scroll fit it into the slot in the orbiting scroll This tip seal is longer than required so must be cut to the correct length There should be a gap of 20 mm between the end of the tip s...
Page 7: ...ghts reserved Page 3 Edwards and the Edwards logo are trademarks of Edwards Limited Replace the tip seals and exhaust valve A726 01 840 Issue F Figure 1 Replacing the tip seals pre serial number 03787...
Page 8: ...dwards Limited 2010 All rights reserved Edwards and the Edwards logo are trademarks of Edwards Limited Replace the tip seals and exhaust valve Figure 2 Replacing the tip seals post serial number 03787...
Page 9: ...imited 2010 All rights reserved Page 5 Edwards and the Edwards logo are trademarks of Edwards Limited Replace the tip seals and exhaust valve A726 01 840 Issue F Figure 3 XDS5 10 pump side view 1 Leve...
Page 10: ...A726 01 840 Issue F Page 6 Edwards Limited 2010 All rights reserved Edwards and the Edwards logo are trademarks of Edwards Limited This page has been intentionally left blank...
Page 11: ...of Vacuum equipment and Components Procedure HS1 If you are returning a vacuum pump note the following If a pump is configured to suit the application make a record of the configuration before return...
Page 12: ...P800 80 000 Issue T edwardsvacuum com...