Resident  PC  software  and  browser  based  application 
can  be  downloaded  for  free  at 

www.edulab.com

  as 

well as a full software user guide.

  

Note:

  Make  sure  not  to  download  and  install  both 

types of software they will conflict on the computer. 
 

Procedure: 

 
1.

 

Install the Edu-Logger software 

2.

 

Connect the 900-213 module to the PC or Mac 

3.

 

Connect  the  salinity  sensor  to  the  900-213 
module (they directly plug together).  Please note 
there is no calibration required for this sensor. 

4.

 

Open the Edu-Logger software. 

5.

 

Once  a  salinity  sensor  logo  appears  on  the  left 
side  of  the  screen  the  probe  has  been 
automatically  identified  and  you  can  begin 
experimentation. 

6.

 

If  the  salinity  sensor  is  not  automatically 
identified, click on the “Search for sensors” icon 
to find the sensor. 

7.

 

Select the “On-line experiment” button; this will 
open a graph below. 

8.

 

Click  on  the  “Module  setup”  button  located  on 
the salinity sensor icon in the module window to 
change the sensor settings if need be. 

9.

 

Click on the "experiment set up" button to change 
the  experiment  settings  if  need  be  (experiment 
duration for example). 

10.

 

The salinity sensor will give a live reading in the 
box to the left of the screen while plugged in. 

11.

 

To run an experiment and collect data click “Run 
experiment”. 

12.

 

To  end  data  collection  early,  click  “Stop 
experiment”. 

 
 

Tablet, smart phone device 

Summary of Contents for 900-238

Page 1: ...Edu Logger Salinity Logger Sensor Guide 900 238 ...

Page 2: ...the box using this guide Hundreds of possible experiments that can be done with the 900 238 sensors are Monitoring of ecological systems chemical analysis and identification environmental health testing water quality testing chemical properties properties of electricity etc The Edu Logger salinity sensor can use any of the following units of measure Percentages The total saline percentage out of t...

Page 3: ...ntly rinse the electrodes off with distilled water after each use and between samples to avoid damage and contamination Handle and store with care to avoid damaging or chipping the electrodes Quick start procedure PC or Mac Computer Materials needed 900 238 Salinity Sensor 900 213 USB Module USB to mini USB cable included with the 900 213 module Your salinity sensor needs to be connected to a 900 ...

Page 4: ...e screen the probe has been automatically identified and you can begin experimentation 6 If the salinity sensor is not automatically identified click on the Search for sensors icon to find the sensor 7 Select the On line experiment button this will open a graph below 8 Click on the Module setup button located on the salinity sensor icon in the module window to change the sensor settings if need be...

Page 5: ...900 215 module to the right side of the 900 268 module 3 Although not required we recommend plugging the 900 215 to an outlet using a USB to mini USB charger such as a typical cell phone charger The 900 268 module will run for 15 75 minutes depending on the sensor without being plugged in 4 The 900 268 can be powered directly using a mini to USB cord and plugging it into your computer or a wall ch...

Page 6: ...uto detect the sensor s select Search for sensors 13 Once the sensor is found you will see an icon on the left side of the screen for the salinity sensor The icon will display real time data 14 Click on the Module setup button located on the salinity sensor icon in the module window to change the salinity probe settings if need be 15 Click on the experiment set up button to change the experiment s...

Page 7: ...ill be monitoring data in real time 5 To run an experiment and collect data click the run button little green person Off line experiments Off line experiments are for when you do not have a sensor connected directly to a computer tablet smartphone or Edu Logger viewer Materials needed 900 238 Salinity Sensor 900 215 Battery Materials needed to configure your offline experiment 900 213 Module or 90...

Page 8: ...device s quick start procedure section 5 Open Edu Logger software 6 Click the Off line experiment button 7 Click the Load data from sensors button 8 Select which experimental data to upload 5 experimental runs can be stored at one time on the 900 238 salinity sensor Restoring sensor s factory default settings Procedure 1 Connect the 900 213 to a PC Mac tablet or smart device 2 Connect the salinity...

Page 9: ... 16 bit 0 0064 100 0 to 64 000 mg L 64 mg L 0 to 64 000 ppm 64 ppm Sensor features Fully digital data Rugged plastic ergonomic case Push button switch for Start Stop experiments in off line mode LED indicator of experiment status blinks while collecting data Pre calibrated sensing equipment Salinity sensing electrodes located on the probe which is attached directly to the sensor s plastic body Not...

Page 10: ... upgraded at any time using software On the salinity probe there are two electrodes of known surface area with a known distance between them which allows for the sensor to determine the salinity of solutions The electrodes located on the salinity probe have an electrical potential difference applied to them to generate a current through the solution If ions are present the current will flow betwee...

Page 11: ...o C 50 o C Never place the probe in viscous heavy oils glycerin ethylene glycol acetone or organic solvents such as pentane or hexane Warranty We promise to deliver our sensor free of defects in materials and workmanship for a period of 3 years from the date of purchase Our warranty does not cover damage of the product caused by improper use abuse or incorrect storage Sensors with a shelf life suc...