EDT directION QC355TX User Manual Download Page 4

Section 2 -

 

pH Calibration and Measurement

For accurate results, stir all buffers and samples. Always rinse electrodes and blot dry 
before transferring from one solution to another to prevent contamination. Ensure that any 
electrode filling holes are left uncovered during use.

Automatic  Calibration & Temperature Compensation

1. 

Connect the pH electrode(s) and temperature probe and switch on using the power 

ON/OFF switch on the back panel.

2.

 Select pH mode by pressing  the pH/°C key. Note:  A flashing display indicates that the 

instrument is uncalibrated.

3. 

Place the pH electrode(s) and the temperature probe in the pH 7 buffer and wait for a 

stable reading. Indicated by 

 on the display.

4.

 Press the CAL key– CAL flag will flash. Press the 7pH key—CAL flag will stop flashing. The 

first calibration point has now been entered automatically at the correct value for the 
temperature of measurement.

5. 

Rinse, blot and place electrode(s) and temperature probe in pH buffer 4 or 10. Wait for a 

stable reading indicated by 

on the display. 

6.

 Press the CAL key– CAL flag will flash. Press the 4/10pH key—CAL flag will stop flashing. 

The second calibration point has now been entered automatically at the correct 
value for the temperature of measurement.

7.

 pH measurements may now be made by immersing rinsed electrodes in the sample and 

recording the stable reading indicated by the 

 flag.

8.

 To read the temperature press the pH/°C key at any time.

7pH

4/10pH

Cal

Send

pH/°C

Calibration Keys, automatically 
selects and enters correct pH value

Starts calibration routine

Selects pH or temperature mode

Transmits data to printer or computer 

Figure 3. Front Panel

Summary of Contents for QC355TX

Page 1: ...on 1 Introduction and Set Up Section 2 pH Calibration Measurement Section 3 Use with a Printer or Computer Section 4 Section 5 Appendix 1 pH Electrodes Troubleshooting Specification QC355TX Routine QC...

Page 2: ...their examination After examining and removing contents save packing material in the event that re shipment be necessary The following items are packed in the box QC355TX Meter E8051 Temperature Probe...

Page 3: ...ell types Combination or sensing half cells should have a BNC terminal and be connected to the socket marked pH ION input at the back of the meter Reference half cells should have a 4mm bunched termin...

Page 4: ...ash Press the 7pH key CAL flag will stop flashing The first calibration point has now been entered automatically at the correct value for the temperature of measurement 5 Rinse blot and place electrod...

Page 5: ...inated with a Carriage Return CR Line Feed LF All characters are ASCII printable alpha numeric Three commands CA PR and RD can be sent from the computer DATE OPERATOR SAMPLE pH 7 00pH T 21 7 C Command...

Page 6: ...meter for servicing Error Codes PRO Temperature probe malfunctioning or not connected Buf Wrong or Contaminated Buffer SL Poor Electrode Slope caused by faulty electrode or poor buffers standards E Fa...

Page 7: ...soaking in pH storage solution for 30 minutes Storage Put some pH storage solution in the protective cap and place over the glass bulb Never store the electrode in distilled or de ionised water Never...

Page 8: ...e information on our products please visit our website www edt co uk EDT directION Unit 5 Waldershare Park Dover CT15 5DQ Telephone 01304 829960 Email sales edt co uk Company No 04135318 VAT No GB 765...