EDT directION GP353 User Manual Download Page 4

Section 2 -

 

pH Calibration and Measurement

For accurate results, stir all buffers and samples. Always rinse electrodes and blot dry 
before transferring from one solution to another to prevent contamination. Ensure that any 
electrode filling holes are left uncovered during use.

Automatic Temperature Compensation

1.

 Connect the pH electrode(s) and temperature probe and turn the mode switch to °C

2.

 Place the pH electrode(s) and the temperature probe in the first buffer, and note the 

temperature of the solution. Refer  to the temperature chart on the buffer  (or see 
Appendix 1) and note the correct value of the buffer at the noted temperature.

3.

 Turn the mode switch to pH mode. Wait for the reading to  stabilise. Using the CAL 

control, adjust the reading to the value obtained in step 2.

4.

 Place the pH electrode(s)  and temperature probe in the second buffer. Again establish 

the correct value for the temperature of measurement. When the reading is stable, 
adjust to the correct value using the Slope control.

5.

 Automatically temperature compensated readings may now be made by immersing the 

pH electrode(s) and temperature  probe into the sample and recording the stabilised 
reading.

Manual Temperature Compensation

If use of a temperature probe is inappropriate (e.g. small sample size), then it is possible to 
use manual temperature compensation).

1. 

Disconnect temperature probe.

2. 

Measure the temperature of the buffer with a thermometer. 

3. 

Switch to °C mode and adjust the reading to the correct value using the Temp °C control.

4.

 Refer to the temperature chart on the buffer (or see Appendix 1) to establish the correct 

buffer value for the temperature of measurement.

5.

 Follow the procedure given above for Automatic Temperature Compensation,  points 3. 

to 5. above, measuring and adjusting the temperature manually for each buffer and 
sample.

No Temperature Compensation

For accurate results with no temperature compensation, the temperature of all buffers and 
samples should be equal and remain constant.

1.

 With the temperature probe disconnected, switch the instrument to °C mode and set the 

displayed reading to the temperature of measurement using the Temp °C control.

2.

 Switch to pH mode, place the electrode(s)in the first buffer, wait for the reading to 

stabilise and enter the correct value (given on the buffer packet or in Appendix 1) using the 
CAL control.

3.

 Place the electrode(s) in the second buffer. When the reading is stable, adjust to the 

correct value using the Slope control.

4.

 Place the electrode(s) in the sample and record the stabilised reading.

Summary of Contents for GP353

Page 1: ... Set Up Section 2 pH Calibration Measurement Section 3 mV Calibration Measurement Section 4 Use of the Recorder Output Section 5 Troubleshooting Section 6 Specification Appendix 1 pH Electrodes Buffers GP353 General Purpose pH Meter Users Manual ...

Page 2: ...mining and removing contents save packing material in the event that re shipment be necessary The following items are packed in the box GP353 Meter E8050 Temperature Probe E8060 Electrode Stand AC Adapter 9V Battery pH Buffers 4 7 10 Operator s Manual Setting Up AC Operation Only use the approved power adaptor supplied Check that the adaptor is the correct voltage for your power supply Plug the ad...

Page 3: ...ready for calibration A Note on Electrodes Combination pH electrodes reference and sensing electrode in one body or two half cells reference and sensing electrodes separate may be used with the GP353 Combination or pH half cells should have a BNC terminal and be connected to the socket marked pH input at the back of the meter Reference half cells should have a 4mm bunched terminal and be connected...

Page 4: ...ing Manual Temperature Compensation If use of a temperature probe is inappropriate e g small sample size then it is possible to use manual temperature compensation 1 Disconnect temperature probe 2 Measure the temperature of the buffer with a thermometer 3 Switch to C mode and adjust the reading to the correct value using the Temp C control 4 Refer to the temperature chart on the buffer or see Appe...

Page 5: ... 60mV 4 mV values relative to the standard solution may now be taken by immersing in the sample and recording the reading Use of the recorder output 1 Connect the recorder via the red and black 4mm sockets on the back panel Red positive Black negative 2 Ensure that the recorder is set for the appropriate range as shown below MODE RANGE mV DISPLAY RECORDER pH 0 200 7 00pH 70 0mV mV 200 500mV 50 0mV...

Page 6: ... Reference filling solution contaminated Buffers contaminated Erratic drifting readings or display reads on left hand side when shorting plug is attached Return meter for servicing pH Range 2 to 19 99 pH Resolution 0 01pH pH Accuracy 0 02pH Temperature Compensation 0 100 C Manual Calibration 2 points at any value mV Range 2000mV mV Resolution 1mV mV Accuracy 2 1 digit C Range 100 to 200 C C Resolu...

Page 7: ...electrode storage solution is made up by dissolving 1g KCl in 100mL of pH 7 buffer pH Buffers Buffer Capsules Buffer capsules are made up as follows For each buffer empty the powder into a suitable container Using deionised water make up to 100mL ensuring that the powder is filly dissolved before use The coloured outer skin may be added to colour code the resulting buffer This may take up to 4 hou...

Page 8: ...re information on our products please visit our website www edt co uk EDT directION Unit 5 Waldershare Park Dover CT15 5DQ Telephone 01304 829960 Email sales edt co uk Company No 04135318 VAT No GB 765175410 ...

Reviews: