Appendix 1 -

 

pH electrodes

Before Use 

Remove the protective cap covering the glass sensing bulb and replace with the protective guard if 
applicable. Inspect the filling solution for air bubbles and remove by shaking in a downward direction. 
Soak the electrode in pH storage solution for 30 minutes.

Cleaning

Soak the electrode in 0.1M HCl for 15 minutes followed by soaking in pH storage solution for 30 
minutes

Storage.

Put some pH storage solution in the protective cap and place over the glass bulb.

Never store the electrode in distilled or de-ionised water. Never allow the electrode to dry out.

pH electrode storage solution is made up by dissolving 1g KCl in 100mL of pH 7 buffer

Buffer Capsules

Buffer capsules are made up as follows For each buffer, empty the powder into a suitable container. 
Using deionised water, make up to 100mL ensuring that the powder is filly dissolved before use. The 
coloured outer skin may be added to colour code the resulting buffer. This may take up to 4 hours to 
dissolve but will not affect the pH of the buffer.

Buffer Solutions

If using ready made solutions, use ones supplied  with  temperature coefficient information

Table of Temperature Coefficients For EDT Buffers

pH Buffers

°C

 

pH 4

 

pH 7

 

pH 10

10

3.99

7.07

10.18

15

4.00

7.04

10.14

20

4.00

7.02

10.06

30

4.00

6.99

9.95

35

4.01

6.98

9.91

40

4.02

6.97

9.85

50

4.05

6.96

9.78

60 

4.07 

6.96 

9.75 

25

4.00

7.00

10.00

 

Summary of Contents for FE257

Page 1: ...et Up Section 2 pH Calibration Measurement Section 3 mV Calibration Measurement Section 4 Use of the Recorder Output Section 5 Troubleshooting Appendix 1 pH Electrodes Buffers Section 6 Specification...

Page 2: ...t Note The carrier will not honour any claims unless all shipping material is saved for their examination After examining and removing contents save packing material in the event that re shipment be n...

Page 3: ...ue in the range 0 0 to 19 0 C and is adjustable i e responds to the use of the keys Adjust to 20 0 C 6 The meter is ready for calibration A Note on Electrodes pH Ion Selective and Redox electrodes may...

Page 4: ...ust the E value of the electrode place in buffer 4 7 or 10 and press the BUF 1 key Note The performance of the electrode may be monitored by pressing the SLOPE key The slope is displayed as a of Nerns...

Page 5: ...i e Section 3 mV Calibration and Measurement The mV mode has two ranges The meter will automatically select the most appropriate range as follows 0 to 400mV with resolution 0 1mV 400mV to 2000mV with...

Page 6: ...n left hand digit Over range condition Drifting readings Erratic drifting readings or display reads on left hand side when shorting plug is attached Section 5 Troubleshooting Probable Cause Battery is...

Page 7: ...electrode storage solution is made up by dissolving 1g KCl in 100mL of pH 7 buffer Buffer Capsules Buffer capsules are made up as follows For each buffer empty the powder into a suitable container Usi...

Page 8: ...e information on our products please visit our website www edt co uk EDT directION Unit 5 Waldershare Park Dover CT15 5DQ Telephone 01304 829960 Email sales edt co uk Company No 04135318 VAT No GB 765...

Reviews: