TEMPERATURE CALIBRATION
AIR MOVEMENT WILL AFFECT THE TEMPERATURE READING.
WORK IN AN AREA WHERE THIS IS MINIMUM.
1. Use the
DS299S
to remove the temperature control knob. Turn Temp. Control
(A)
so arrow
points to cord.
2. Use the
SRW01
to pull out heater assembly, PCB & power cord from handle. Leave
enough power cord slack to install
SRF30
.
3. Place heater assembly, PCB & power cord inside
SRF30
. Install tip that is clean and well
tinned.
4. With the use of the
MS412 Temperature Caliration System
. Follow set-up shown below.
5. Plug in tool
and by using the Pot Adjustment Tool supplied, turn Temp. Control
(A)
to
"solder-melt" temperature.
6. Apply a small amount of solder on the Tip, just enough to form a solder bead on top of
the Tip. Place the center of the thermocouple wire of the on top of the solder bead.
7. Again, apply a small amount of solder on the center of the cross wire, just enough to
embed the center.
8. Turn Temp. Control
(A)
fully clockwise. Adjust Calibration Pot
(B)
until reading stabilizes at
800°F.
9. Assemble unit in original handle.
10. Before installing the Temp. Control Knob, peel-off pointer label from the Knob.
11. Using same MS412 set-up, install and adjust Temp. Control Knob so Meter will read desired
temperature.
12. Attach new pointer label in accordance with the reading of the Meter.
Calibration Pot
(B)
SRF30
original handle
Temp. Control
(A)
Tools Needed
•SRW01
•DS299S
•SRF30
•MS412
•Pot Adjustment Tool
SRW01
DS299S
MS412 Temperature
Calibration System
SDS100
SRF30
FX635
TI680 Meter
Pot Adjustment Tool
POINTER LABEL