4

Detachable optical encoder

How to install and remove from scanning head

The encoder is fastened to the Pipescan on the side of the electronic module.

Install the encoder prior to deploying the head on the part to inspect to have easy access to the slot.

To do so, position the encoder’s fixation bracket in its slot and fasten using the two allen screws (1), as shown.

Connect the encoder’s connector. Make sure that red dots on connector and receptacle are aligned when doing so.

To remove, simply disconnect the connector and remove the two allen screws.

Cable

Maintenance of Scanning Heads

Magnets attract dirt. When deemed necessary during inspection or after an inspection, clean the reading surface and 
magnet with soft cloth. 

Do not immerse Pipescan HD.

Always follow general precautions and safety precautions during maintenance of the probe.

Wheels can also be replaced if required. If needed, spare parts can be ordered directly from Eddyfi Technologies.

Contact wheel

Pivoting arm

Fixation bracket 
(Attach to pipescan HD 
with 2 screws M3)

4-pin Lemo connector

Fixing screws

41-pin Amphenol connector. Head side.

I/O connector 

Signal connector (160-pin array)

Drive connector 

Summary of Contents for PIPESCAN HD

Page 1: ...D SWIFT M MFL Data Acquisition Instrument 1 What s In The Box 2 2 Swift M Overview 2 3 Pipescan HD Overview 3 4 Software Overview 5 5 Pipescan Setup 7 6 Work Flow Software 8 7 Analysis and Reporting 1...

Page 2: ...USB 2 0 connectors Network connector HDMI connector Audio connector Quick Copy Button MFL Connector I O connector 12 pin Fischer Drive connector 19 pin Fischer 160 pin array connector 1 WHAT S IN THE...

Page 3: ...escan head should be adjusted to nominal OD while the head is not on the inspection surface Adjusting the curvature while on the part could damage the mechanism due to the strength of the magnets bein...

Page 4: ...so To remove simply disconnect the connector and remove the two allen screws Cable Maintenance of Scanning Heads Magnets attract dirt When deemed necessary during inspection or after an inspection cl...

Page 5: ...ve all indications above threshold to report 5 No defect Classifying an indication as not a defects 6 Review N A 7 Add Note Add general note to C Scan 8 Capture Screen shot of current view Layout 1 Re...

Page 6: ...to be automatically identified 8 Show Palette controls Display the colour pallet settings 9 Move Cursor Change of cursor position and size 10 X axis N A Current View Stripchart 1 Indications Allows in...

Page 7: ...he system 1 2 Adjustment using a ruler If you can t find the gauge or if the diameter needed is not on it use a ruler and adjust the system The measurement must be taken between the outer posts on eac...

Page 8: ...etails 2 1 Project Folder 2 2 Inspection 3 Select Create Setup 3 1 Select Probe 3 2 Configure Scan definition 4 When the Probe is ready to perform the first Setup scan scan press the Acquire button th...

Page 9: ...an alarm in the strip chart meaning any indications above this alarm are automatically tagged as indications of interest To set the alarm i Select the strip chart user will see a purple perimeter aro...

Page 10: ...n x Filename indication from all files of the inspection folder can be added in the same report x Index Numerical increment for each indication added to the report automatically generated by the softw...

Page 11: ...101 6 0 276 7 0 4 0 4 5 114 3 0 189 4 8 PS HD MFL 127XFLAT 4 5 5 0 127 0 1 15 29 2 5 5 563 141 30 1 067 27 1 6 6 625 168 28 0 937 23 8 7 7 625 193 68 0 839 21 3 8 8 625 219 08 0 76 19 3 9 9 625 244 4...

Reviews: