MU300 Rev1.0 Oct20
Page 36 of 54
Select the transducer installation methods, the
following methods are available:
0.V
1.Z
2.N
Modes
Mode menu is defined by the manufacturer.
Please don’t make any alterations.
14
Installation Spacing
Apply the installation spacing of transducers
calculated by the flowmeter according to pipe
parameters you set.
20
Damping
Damping coefficient is 0~999 sec.
0: Indicates no damping; 999: Indicates the
maximum damping.
Damping is used for smoothing the displayed
data.In general, you need to enter one digit
among 3~10.
21
Low Vel. Cut off
Low Flow Cut off is used to make the system
display as "0" value at lower and smaller flows
to avoid any invalid totalizing. For example, if
the cutoff value is set as 0.03, system will take
all the measured flow velocity values from -
0.03 to + 0.03 as "0". Generally, 0.03 is
recommended in most applications
22
13
Transducer
*R
Size
Method
Mode
Options
0.V
13
Transducer
*R
Size
Method
Mode
Options 0.Mode o
Other
0.000
14
INST Space
*R
Value
63.2
mm
20
Damping
*R
Value
6
21
Low Vel. Cut off
*R
Value
0.03
m/s