Installation Requirements
Element Basic User Guide
Page 11 of 21
3
INSTALLATION AND ENVIRONMENT REQUIREMENTS
3.1
POWER REQUIREMENTS
Component
Rating
Workstation Power
550 watts (maximum); 100-240 Volts
Monitor
55 watts (typical) ; 100-240 Volts
Detector
20 watts (max) (< 10 w typ.) ; 100-240 Volts
Table 1 - Power requirements
3.2
SPACE AND WEIGHT SPECIFICATIONS
Workstation dimensions:
17.7 x 6.7 x 18.0 in (44.9 x 17.0 x 45.7 cm)
Workstation weight:
33.0 lb. (14.97 kg)
Detector Body dimensions
1.75 (W) x 4.0 (H) x 3.25 in (L) (45 x 105 x 82 mm)
Detector Weight
5 lb. (2.2 kg)
3.3
ENVIRONMENT REQUIREMENTS
Parameters
Operating Conditions
Temperature
0
C to 35
C
Humidity
0% to 85% RH, non-condensing
Barometric Pressure
525 to 800 mm of Hg
Air Velocity
0 to 0.5 m/sec
Table 2 - Environment requirements / Operating conditions