Man down alarm

Falling by breathing dangerous gases can cause serious injury and even fatality to workers. If the 
Man  down  alarm  function  in 

ES

SAN-10  is  set  on, 

ES

SAN-10  can  detect  falls  and  send  a  man 

down alert which will activate the audible and visual alarms and alert other people in the area. 

The  man-down  detection  uses  a  three-axi

s  accelerometer  to  automatically  monitor  the  user’s 

movements.  If  a  sudden  fall/impact  is  followed  by  a  lack    of    movement    for    6    seconds,    the 
alarm will sound. Turn off the alarm by pressing either of the two buttons. 

See Section 2.2 Menu and Configuration Settings, to change settings. 

7. Reset Button

Users can reset the unit by pushing a reset button through a
hole on back of shell.

MAINTENANCE

 

Calibration

 

The 

ES

SAN-10  comes  pre-calibrated  from  the  factory.  However,  the  CO2  sensor  should  be 

calibrated at least once a  year, 

or as described in your company’s 

safety procedures. You can 

perform the calibration yourself, or you can return it to 

us

 for factory calibration at a nominal 

fee. 

The temperature and humidity sensors do not require calibration and should remain accurate for 
the life of your unit. 

Cleaning and Storage

 

The  front  cover  and  case  can  be  cleaned  with  a  mild  solution  of  detergent  and  water. 
Apply sparingly  with  a  soft  cloth  and  allow  to  dry  completely  before  using.  Do  not  use 
aromatic hydrocarbons or chlorinated solvents for cleaning. 

Summary of Contents for ESSAN-10

Page 1: ...User Manual R1 0 ESSAN 10 Pocket CO2 Meter Distributed Supported By Edaphic Scientific Pty Ltd www edaphic com au info edaphic com au Ph 1300 430 928 ...

Page 2: ...ts INTRODUCTION 3 FEATURES 3 MONITOR 3 LCD DISPLAY 4 OPERATION 4 MENU CONFIGURATION 5 Calibration 6 Alarm Threshold 6 MAN DOWN FEATURE 7 MAINTENANCE 7 SPECIFICATIONS 8 TROUBLESHOOTING 9 WARRANTY 9 CONTACT US 9 ...

Page 3: ...udible visual strobe and vibrating alarms Large LCD display Man down alarm leveraging accelerometer technology Rugged design with protective rubber enclosure Heavy duty metal clip Front facing sensor unit Rechargeable Li ion Battery 4 2v 1500mAh 18 hours percharge Black Micro USB cable and Wall USB charger Dual calibration methods Ambient Air 300 600ppm and Nitrogen 0 200ppm Automatic atmospheric ...

Page 4: ...n the monitor off or change the C F setting 1 Power button 1 1 When the Monitor is turned off press to turn on the unit 1 2 When the Monitor is turned on press for 3 seconds to turn off theunit When the unit is first turned on it performs 5 seconds countdown for Monitor warm up then enters normal display with current CO2 temperature and humidity readings displayed The monitor starts taking measure...

Page 5: ...sesthe power button to implement Nitrogen Calibration 0 200ppm H0 5 H0 5T H1 0 H1 5 H3 0 Current Defaults are L 0 5 H1 0 High Alarm Threshold setup User presses thepower button to switch the high alarm threshold H0 5 5000 ppm H0 5A 5000 ppm H1 0 10000 ppm H1 5 15000 ppm and H3 0 30000 ppm L0 5 L0 5A L1 0 L1 5 L3 0 Low Alarm Threshold setup User presses the power button to switch the low alarm thre...

Page 6: ...ure Limit high alarm The alarm thresholds can be changed dependent on the user to L0 5 5000 ppm L0 5T 5000 ppm TWA L1 0 10000 ppm L1 5 15000 ppm or L3 0 30000 ppm If the measured CO2 value exceeds the define Low Alarm threshold the alarm LED will flash red at 2Hz and the audio alarm de de de de will sound If the CO2 level exceeds the High Alarm threshold the LED will flash faster and the audio ala...

Page 7: ...ettings to change settings 7 Reset Button Users can reset the unit by pushing a reset button through a hole on back of shell MAINTENANCE Calibration The ESSAN 10 comes pre calibrated from the factory However the CO2 sensor should be calibrated at least once a year or as described in your company s safety procedures You can perform the calibration yourself or you can return it to us for factory cal...

Page 8: ... Pressure Dependence 0 13 of reading per mmHg Response Time About 2 min for 90 of step change Warm up Time 5 seconds at 22 F IAQ LED Display Green 800ppm Yellow 800 1200ppm Red 1200ppm Measurement interval 2 seconds Temperature Sensor Specifications Temperature Range 50 140 F 10 60 C display Display Resolution 0 1 F 0 1 C Display Options F C switchable Accuracy 0 9 F 0 5 C Response Time 5 30 secon...

Page 9: ... turn on contact support Slow response Power cycle Verify Unit Readings do not change Confirm the monitor has sufficient charge WARRANTY The monitor comes with a one 1 year warranty starting from the date the monitor was shipped to the buyer CONTACT US Distributed Supported By Edaphic Scientific Pty Ltd www edaphic com au info edaphic com au Ph 1300 430 928 ...

Reviews: