VE-H100B Veterinary Pulse Oximeter User Manual
- 97 -
and TV broadcast cannot be predicted theoretically with
accuracy. To assess the electromagnetic environment due to
fixed RF transmitters, an electromagnetic site survey should
be considered. If the measured field strength in the location
in which the oximeter is used exceeds the applicable RF
compliance level above, the oximeter should be observed to
verify normal operation. If abnormal performance is
observed, additional measures may be necessary, such as
reorienting or relocating the oximeter.
A2.4 Recommended Separation Distances
For pulse oximeter and charger stand:
Recommended separation distances between portable
and mobile RF communications equipment and the
VE-H100B and charger stand
VE-H100B and charger stand is intended for use in an
electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of
VE-H100B
and
charger
stand
can
help
prevent
electromagnetic interference by maintaining a minimum
distance between portable and mobile RF communications
equipment (transmitters) and the oximeter as recommended
below, according to the maximum output power of the